Oven |
- Maximum capacity: 22 L
- Temperature range: RT + 5 ℃ to 400 ℃
- Temperature accuracy: ± 0.1 ℃
- Temperature program: 9-phase/ 10 platform
- Program total time: 9999.9 min
- Maximum temperature heating rate: 0.1 to 60 ℃/ min
- Maximum cooling rate: ˂ 10 mins (250 to 50 ℃)
|
Sample injector |
- Temperature range: RT+ 7 ℃ to 420 ℃
- Temperature control: Automatic
- Injector type: Column or split
- Maximum sample inlet: 3 pcs
- Carrier gas flow type: Constant pressure
- Pre column pressure range: 0 to 400 kPa
- Pre column pressure accuracy: 0.1 kPa
- Mobile phase flow range: H2: 0 to 200 ml/min, N2: 0 to 150 ml/min
|
Detector |
- Temperature range: RT + 7 ℃ to 420 ℃
- Maximum installation: Dual detector
- Flame ignition mode: Automatic
- Flame Ionization Detector (FID): Logarithmic amplifier <br/>
High voltage switch control <br/>
Baseline signal display <br/>
Ignition coil control <br/>
Detection limit: ≤3 ×10 to 12 g/s
(Sample: n-hexadecane) <br/>
Baseline noise: ≤ 5 × 10 <sup> - 14 </sup> A <br/>
Baseline draft: ≤ 6 × 10 <sup> -13 </sup> A <br/>
Dynamic range: 10 <sup> 7 </sup> <br/>
RSD: ≤ 3 %
|
Power supply |
AC 220 V ± 22 V, 50 Hz ± 0.5 Hz, 3000 W |
Packing dimension |
1020 × 660 × 740 mm |
Gross Weight |
82 kg |
Net weight |
70 kg |
- 7-inch LCD display specifies temperature, gas flow and heating curve
- H2 and N2 mobile gas phase with AFC and constant pressure gas flow type
- Equipped with capillary analysis system and capillary split sampler
- Injection system incorporates diaphragm cleaning, split flow and auxiliary gas adjustment
- Standard unit for high-speed analysis with polar capillary column
- PEG-20M and FFAP stationary phase polar capillary columns can be coupled
- Dual FID detector with automatic flame ignition and baseline signal display
- Self-diagnostic and automatic fault recognition function
- Multi-core and 32-bit embedded hardware system ensures the reliable operation
- Equipped with chromatography workstation software to process data
- Memory function can store 20 sample test data
- RS232 interface and LAN port
Accessories no |
Accessories name |
1 |
Polar capillary columns PEG-20M
Stationary phase: 30 m × 0.32 mm
injector temperature: 250 ℃
Oven temperature: 165 ℃
Detector: FID- 250 ℃
Inlet pressure: 0.04 Mpa
|
2 |
Polar capillary columns FFAP
Stationary phase: 30 m × 0.32 mm
Injector temperature: 250 ℃
Oven temperature: 165 ℃
Detector: FID- 250 ℃
Inlet pressure: 0.04 Mpa
|
3 |
Syringe 1 ul |
4 |
Syringe 5 ul |
Suitable for analysis of trace toxic compounds, monitoring and research in environmental protection, atmosphere and water source pollution. It is widely used for petrol-chemistry, environmental protection, epidemic prevention, pharmacology and scientific research.