Elemental Analyzer

Elemental (CHN) Analyzer
Elemental (CHN) Analyzer LCHN-A10
Elemental (CHN) Analyzer LCHN-A10
Gas required Helium: 99.99% 0.25 ± 0.01 mPa
Oxygen: 99.5% 0.25 ± 0.01 mPa
Nitrogen or Composed air: 0.25 ± 0.01 mPa
Measurement range Carbon: 0.02 to 150 mg
Hydrogen: 0.1 to 12 mg
Nitrogen: 0.04 to 50 mg
Consumption Helium 200 ml/ min
Sample mass Up to 1000 mg, depending on the sample matrix
Analysis time 4 to 6 mins, depending on sample combustion condition