Semi-automatic Bio-chemistry Analyzer LSBA-A20

Semi-automatic Bio-chemistry Analyzer LSBA-A20 also known as chemistry analyzer is a compact, semi-automated and high performance analyzer comprises 8 assay modes offers photometric range from 0 to +3.0 OD with 3 level temperature range along with temperature accuracy of ±0.1˚C. Incorporated with excellent QC function with large memory. Equipped with different wavelengths such as 340 nm, 405 nm, 492 nm, 510 nm, 546 nm, 578 nm, 620 nm with accuracy of ±2nm. Built-in thermal printer with Real time graph display.

Measuring System Measuring Range: 0 to 3.0 OD<br/> Method: Absorbance, End Point, Fixed Time, Rate, Sample Blank, Multi Standard, Two point Non Linear, and Sample Blank Non Linear<br/> Photometric linearity: ±2% (0 to 3.000 OD)<br/> Photometric accuracy: ±1% (0 to 3.000 OD)<br/> Carry over: ≤1%
Photometric System Light source: 6 V, 10 W halogen lamp; standby function<br/> Filters: 340 nm, 405 nm, 492 nm, 510 nm, 546 nm, 578 nm, 620 nm and 2 more open filter positions (on request)<br/> Wavelength Accuracy: ±2 nm
Thermostatic Control Temperature: 25˚C, 30˚C, 37˚C<br/> Precision: ±0.1˚C
Drift <0.005 OD/hr
Reading Cuvette Cell and Direct reading cuvette
Measuring Volume 200 µl to 800 µl
Incubating positions 20 incubation positions
Test Menu offered T.BILI, r-GT/GGT, Ca, APOB-1/B, CHE, D.BILI, ALP/AKP, TG, CL, APOB, CREA, TP, UREA, CK, P, CK-MB, Mg, ALB, LDH, C, ASO, AFU, Lactic Acid, ALT/GPT, UA, α-HBDH, CO2, FMN, LP(a), AST/GOT, GLU, AMY, LDL-C, HDL-C, LA, TBA, ADA, HS-CRP, PA, Fe, Cu, Zn, C3, C4, IgG, IgM, IgA and drug, toxicity etc.
Quality Control QC function with displaying and printing
Display Back illuminated color touch panel; LCD display
Storage Over 200 programs, 1000 test results
Printer Built-in thermal printer
Interface RS-232, USB
Power Supply AC 110 V/ 60 Hz or 220 V/50 Hz
Dimension (W×D×H) 340×380×180 mm
Weight 8 kg
  • Two test mode: flow-cell or cuvette
  • Supports bi-chromatic tests for end point, fixed-time, kinetics methods
  • External keyboard or mouse support via USB
  • User-friendly interface and multi-language operation software
  • 7 inch back illuminated color touch panel; LCD display with in-built thermal printer
  • 20 incubating positions
  • Equipped with both Monochromatic & Bi-chromatic measurement with 6 filters and 2 free positions
  • Halogen lamp as light source with long life span
  • Easy to use with low maintenance

Semi-automatic Bio-chemistry Analyzer is used across molecular biology, biotechnology, immunology, microbiology, routine biochemistry, hormonal assay, electrolytes, therapeutic drugs and drug-enzyme investigations.

Absorbance Range 0 to 3.0 Abs
Aspiration Volume 100 to 5000 µl
Wavelengths 340 nm, 380 nm, 405 nm, 505 nm, 546 nm, 578 nm, 620 nm and 1 free position
Principle Colorimetery, turbimetry
Absorbency Range -0.3 to 3.0 abs
Resolution 0.001 abs (display), 0.0001 abs (inner calculation)
Temperature Accuracy ± 0.1°C
Wavelength 340 nm, 405 nm, 510 nm, 546 nm, 578 nm, 620 nm + 2 free positions
Measuring System Measuring Range: 0 to 3.0 OD<br/> Method: Absorbance, End Point, Fixed Time, Rate, Sample Blank, Multi Standard, Two point Non Linear, and Sample Blank Non Linear<br/> Photometric linearity: ±2% (0 to 3.000 OD)<br/> Photometric accuracy: ±1% (0 to 3.000 OD)<br/> Carry over: ≤1%
Photometric System Light source: 6 V, 10 W halogen lamp; standby function<br/> Filters: 340 nm, 405 nm, 492 nm, 510 nm, 546 nm, 578 nm, 620 nm and 2 more open filter positions (on request)<br/> Wavelength Accuracy: ±2 nm
Thermostatic Control Temperature: 25˚C, 30˚C, 37˚C<br/> Precision: ±0.1˚C
Drift <0.005 OD/hr