HPLC system LHLC-B10 Catalog

Overview

This HPLC system is incorporated with advanced high pressure injection technique. The detector is equipped with 24 bits AD transducer to assure low noise, low drift and high sensibility. This HPLC system can conveniently change the sample input volume. It comes with dual piston reciprocating and series pump for consistent and stable flow.

Specifications
High pressure pump High pressure gradient pump
Peak operating pressure 45 MPa, Stainless steel pump
Pressure ± 0.02 MPa
Flow rate range 0.001 - 9.999 mL / min
Flow rate precision ± 5 %
Flow rate repeatability RSD ≤ 0.1 %
Power 220V ± 10 %, 50 Hz, 250 W
Size (W × H × D) 520 × 420 × 160 mm
UV Detector
  • Wavelength: 190 - 680 nm
  • Wavelength error: ≤ ± 1 nm
  • Baseline noise: ≤ ± 2 × 10-5 AU (Methyl alcohol, 1 mL / min, 254 nm, 20 ºC)
  • Wavelength repeatability: 0.2 nm
  • Flow cell: 8 µL
  • Drift: ≤ 5 × 10-4 AU/h (Methyl alcohol, 1 mL / min, 254 nm, 20 ºC)
  • Limit of detection: ≤ 2 × 10-9 g / mL (Naphthalene methyl alcohol solvent)
Automatic Injector
  • Sample value: Maximum 2 mL, 6 × 25 = 150 units
  • Sample volume: 0 - 1mL, adjustable
  • Sample residual: < 0.02 %
Column oven
  • Temperature control: Room temperature ± 5 ºC - 80 ºC
  • Constant temperature accuracy: ± 1 ºC
  • Constant temperature precision: ≤ 0.1 ºC
  • Temperature setting resolution ratio: 0.1 ºC
Overall Instrument
  • Quantitative analysis: RDS 6 ≤ 0.2 %
  • Quantitative repeatability: RDS 6 ≤ 0.2 %
  • Sample residual: ≤ 0.01 %
  • Gradient error: ≤ 0.01 %

Features

  • Inclusion of high performance patent gradient mixer
  • Intelligent anti-control software with powerful control function
  • Stand-alone TFT colour screen with touch control
  • High pressure gradient pump
Applications :

This gradient HPLC system has multiple applications in medicine analysis, detection of protein content of food, environmental analysis, oil and chemical analysis and in scientific research studies.