Nucleic acid purification system LNAP-A10 is a benchtop unit for DNA and RNA purification with automated low throughput sample preparation Capable of extracting 1 to 20 samples per run with a working volume of 1 ml. Features a fast extraction protocol of 15 to 40 mins per cycle depending on sample type. Built-in UV lamp prevents cross contamination. Unit can be configured for elevation of sample processing volume upto 3 ml and 5 ml for higher yields of purified products.
| Throughput | 1 ~ 20 |
| Process volume | 50 µl ~ 3000 µl |
| Collection efficiency | > 95 % |
| Magnetic rod number | 20 |
| Purification accuracy | 100 copy sample positive rate > 95% |
| Stability | CV < 5% |
| Plate types | 3 ml tube strip |
| Heating for lysis tube | Ambient temperature ~ 120°C |
| Heating for elution tube | Ambient temperature ~ 120°C |
| Operation | 7 inch color touch screen |
| Extraction steps | Lysis, Sample binding, Washing and elution |
| Storage capacity | more than 100 programs |
| Protocol management | Create, edit, delete, protocol mode |
| Pollution control | UV light |
| Lighting | Yes |
| Extension interface | 4 standard USB port, built-in SD card |
| Exhaust | Fan |
| Power supply | 450 W |
| Dimensions | 400 × 520 × 450 mm |
| Weight | 28 kgs |
| Accessory No | Accessory |
| NAP-A1 | Tube strips |
| NAP-A2 | Magnetic rod's tip |
Used for DNA and RNA purification from different starting materials, such as blood, cultured cells or bacteria, tissues, cell-free body fluids and plant samples