Wavelength range |
200 nm to 1000 nm with 1 nm steps |
Wavelength accuracy |
±2 nm |
Wavelength selection |
Monochromator |
Half-bandwidth of filter |
˂ 2.5 nm |
Repetition |
±0.2 nm |
Measurement range |
0 to 4.0 OD |
Detector |
Two silicon photoelectric detector, one for measurement, another for reference |
Linearity at 450 nm |
0 to 2.5 Abs, ±2% (96 well plates) |
Accuracy at 450 nm |
1.0 % + 0.005 Abs (0 - 2.0 Abs), 2.0 % [2.0 to 2.5 Abs) |
Precision at 450 nm |
CV < 0.5 % accurate model, CV < 1.0 % fast mode |
Test speed 01 |
10 seconds with 96-well plate (fast mode), 30 seconds with 96-well plate (accurate model) |
Plate shaking |
Linear, three options for speed |
Test speed 02 |
15 seconds with 96-well plate (single wavelength), 30 seconds with 96-well plate (double wavelength) |
Ambient Temperature |
+5℃ to 45℃ |
Software |
Integrated software or PC control software |
Light source |
Xenon flash lamp |
Display |
10 inch touch screen |
Analysis software |
Reader It- II |
User interface |
Touch screen, android system, 10 inch touch screen, external keyboard mouse |
Storage |
16 G memory, more than 10,000 test records can be stored |
Ports |
3 USB ports, for PC, printer and USB-disk |
Automated systems |
Temporarily unable compatible with automated systems |
Dimension |
300 × 500 × 290 mm |
Power supply |
DC 24 V 6.5 A |
Weight |
15 kg |
- Designed with temperature control incubation system
- Microplate ultraviolet-visible spectrophotometer
- CCD real-time spectral scanning mode and real-time output of UV-visible spectrum
- It measures 96 and 384 well microplates
- Designed with 10 inch touch screen and offers stand-alone operation, it is easy to use
- Built-in grating monochromator
- Wavelength ranges from 200 to 1000 nm
- Microplate oscillating mixing function
- Suitable for cell proliferation and cytotoxicity test
- Designed with flashing xenon lamp source, long life and stable illumination
- USB data interface to facilitate instrument control and data transmission
- It detect the absorbance, concentration and purity of nucleic acids and proteins
- Instrument parameter setting and instrument self-test function is highly automated
- Powerful PC software (Reader It -II), to offer analysis review, validation of data
- It is easy access to data via USB interface and WIFI
Microplate Reader is used in life science research work, especially DNA/RNA analysis, cell proliferation and cytotoxicity test, bacterial concentration analysis, pharmaceutical plant analysis, protease and kinase, phospholipase, NADH, GST activity test and used in end point method ELISA / EIA, with a variety of fitting curves for analysis.