Biochemistry analyzer LBCA-A10 is a fully automated floor standing unit with a direct reading system and random operational access. Features automated washing station for both sample and reagent probe. Provision for liquid level detection a collision protection for both sample and reagent probe. Built-in cooling system for reagent chamber.
| Sample volume | 1 ~ 100 µl , 0.1 µl / step |
| Reagent position | 2 x 40 |
| Reagent volume R1 | 1 ~ 400 µl , 1 µl / step |
| Reagent volume R2 | 1 ~ 400 µl , 1 µl / step |
| Throughput 1 reagent | 400 test/h |
| Throughput 2 reagent | 400 test/h |
| Throughput With ISE | 400 test/h + 120 test/h |
| Sample position | 80 |
| Assay items | 80 assay items |
| Operation | Random access |
| Reading system | Direct reading system |
| Method | End point , Kinetic , Fixed time , Immunoturbidimetric , 1-2 reagent , Multistandard , Reagent / serum blank etc. |
| Reagent / Sample probe | With liquid level sensor , collision sensor , Teflon coating and automatic washing station |
| Reaction time | 0 ~ 999 secs |
| Cuvette | High quality UV transmitted plastic cuvette |
| Washing | 8 channel automatic washing system for reaction cuvettes |
| Calibration | Linear , Non-linear , Multi-standard , K factor |
| Sample dilution / retest | Sample can be diluted and retested if results are out of range or sample is insufficient |
| Wavelength | 340 nm ~ 810 nm |
| Light source | Halogen lamp |
| Absorbance resolution | 0.0001 A |
| Absorbance accuracy | ± 0.0003 A ( 0 ~ 2.5 A ) |
| Repeatability ( CV % ) | ≤ 2 % |
| Quality control | Three level programme |
| Bar code | Optional |
| Printer | Multiple report format |
| Power supply | 220 V ; 50 Hz 110 V ; 60 Hz |
| Dimensions | 1000 x 850 x 700 mm |
| Weight | 137 kgs |
The biochemistry analyzer that we employ is entirely automated and has a direct reading function. It is equipped with a fully automated washing station that may be used for either the sample or the reagent probe. A provision has been included for measuring the level of liquid, and there is also protection against collisions for both the sample probe and the reagent probe. This apparatus is utilised on a regular basis for the determination of the concentration of blood sugar, cholesterol, electrolytes, and calcium in blood, serum, plasma, or urine samples; for the requirement of certain haematology values, such as haemoglobin concentration and prothrombin times, for the diagnosis and treatment of diabetes, thyroid problems, as well as hepatitis and kidney conditions; and for the provision of certain haematology values, such as haemoglobin concentration and prothrombin times, for the determination. In a nutshell, the tool is well-suited for doing essential tasks in an effective and risk-free manner. Developed with alert functions to serve as a deterrent against making mistakes or running into issues. Compatible with the most recent releases of both Windows 7 and Windows 10 operating systems. Highlights ISE module and barcode scanner (optional), Card test control system (on request), Quartz glass cuvettes (on request), Printer etc.