Analytical HPLC System LAHS-B11

Analytical HPLC System LAHS-B11 is a unit equipped with LCD display UV-Visible wavelength range of 190 nm to 700 nm and double plunger reciprocating pump system for constant flow pump. It has gradient kit system for test result accuracy. Consists of separate workstation, sample injector and front mounted self-locking switch. Deuterium lamp is used as light source with valve mounting frame. It is ideally used in the detection of compounds in pharmaceutical industries, research institute, food industries, and petroleum industries, etc.

High Pressure Constant Flow Pump
  • Double plunger reciprocating pump: 2 sets
  • Flow range: 0.001 to 10.00 mL/min
  • Setting step size: 0.001 mL/min
  • Flow accuracy: ≤ ± 0.3 %
  • Flow stability: RSD ≤ 00.1 %
  • Maximum pressure: 42 MPa
  • Display pressure error: ≤ 1.0 %
  • Pressure fluctuation: ≤ 1.0 %
  • Power consumption: 80 W
UV/Visible Detector
  • Wavelength range: 190 to 700 nm
  • Wavelength repeatability: ± 0.1 nm
  • Band width: 8 nm
  • Response time: 0 to 4.9 seconds
  • Linearity range: 1.8 AU (5 %)
  • Baseline Noise: ≤ ± 1 × 10<sup>-5</sup>AU
  • Baseline drift: ≤ 2.0 × 10<sup>-4</sup>AU/h
  • Minimum detectable concentration: 4.0 × 10<sup>-9</sup> g/mL
  • Detector cell: Optical path: 10 mm<hr> Volume: 10 µL
  • Light source: Deuterium lamp
  • Power consumption: 100 W
  • Voltage: AC 110 V/ 220 V ± 10 %, 50/60 Hz
  • Packing Dimensions (W × D × H): 600 × 440 × 380 mm<hr> 3 × (600 × 410 × 300 mm
  • Gross weight: 75 kg
  • Double plunger reciprocating pump system
  • Gradient system kits
  • Sample injector: 1 set
  • Deuterium lamp light source
  • Separate workstation
  • LCD display for value observation
  • Front mounted self-locking switch
  • Output: USB port
  • Compact and easy to install

Widely used in pharmaceutical, petroleum, research institute, food industrial applications in separation and detection of mixture of compounds in chemical analysis, poison analysis, environmental analysis, protein food detection, medicine testing, and polymer analysis.

High Pressure Constant Flow Pump
Double plunger reciprocating pump 1 set
Flow range 0.001 to 10.00 mL/min
Setting step size 0.001 mL/min
Flow accuracy ≤ ± 0.3 %
Flow stability RSD ≤ 00.1 %
Maximum pressure 42 MPa
Display pressure error ≤ 1.0 %
Pressure fluctuation ≤ 1.0 %
Power consumption 80 W
UV/Visible Detector
Wavelength range 190 to 700 nm
Wavelength repeatability ± 0.1 nm
Band width 8 nm
Response time 0 to 4.9 seconds
Linearity range 1.8 AU (5 %)
Baseline Noise ≤ ± 1 × 10<sup>-5</sup>AU
Baseline drift ≤ 2.0 × 10<sup>-4</sup>AU/h
Minimum detectable concentration 4.0 × 10<sup>-9</sup> g/mL
Detector cell Optical path: 10 mm<hr> Volume: 10 µL
Light source Deuterium lamp
Power consumption 100 W
Voltage AC 110 V/ 220 V ± 10 %, 50/60 Hz
Packing Dimensions (W × D × H) 600 × 440 × 380 mm<hr> 2 × (600 × 410 × 300 mm
Gross weight 54 kg
Pump Specifications
Model No HPP-B1
Flow rate range 0.0001 to 10 ml / min
Maximum pressure 42 MPa , 0 to 5 ml / min
Column size 4.6 × 250 mm
Sample size 5 μm
Pressure pulsation ≤ 1.5 % at 10 MPa
Flow rate accuracy ± 0.2 % at 1 ml / min
Flow rate repeatability % ≤ 0.15
Gradient accuracy % ± 2
Gradient repeatability % ≤ 0.2
Dimension 450 × 260 × 190 mm
Power supply 90 V to 240 V
Weight 11.2 kg
Detector Specifications
Model No HPD-B2
Wavelength range 190 to 800 nm
Wavelength accuracy ± 0.75 nm
Wavelength repeatability 0.2 nm
Band width 8 nm
Optical path 10 nm
Deuterium lamp 190 to 400 nm
Tungsten lamp 400 to 800 nm
Baseline noise ( Static ) Au ≤ ± 0.5 × 10 <sup>-5</sup>
Baseline drift ( Static ) Au / h ≤ 1.5 x 10 <sup>-4</sup>
Minimum detection limit ≤ 4 x 10 <sup>-8</sup>
Power 100 W
Power supply 90 V / 50 to 60 Hz
Dimension 460 × 260 × 190 mm
Weight 9.2 kg
Pump specification
Type Quaternary Low Pressure Pump
Flow Rate Range 0.001 to 10.000 ml/min, in 0.001ml increments
Flow Accuracy ≤ ± 0.2% (1.00 ml/min, H<sub>2 </sub>O, R.T., 8.5 MPa)
Flow Precision RSD ≤ 0.06% (1.00 ml/min, H<sub>2 </sub>O, R.T., 8.5MPa)
Maximum Pressure 63 MPa
Pressure accuracy ≤ ± 3.0 %
Pressure pulsation ≤ 0.1 MPa
Composition rang 0 to 100 %
Composition precision ≤ 0.3 % RSD
UV/VIS Detector
Wavelength Range 190 to 800 nm
Wavelength Repeatability ≤ 0.1 nm
Wavelength Accuracy ≤ ± 1.0 nm
Band Width 6 nm
Response Time 0.0 s to 5 s
Baseline Noise ≤ ± 0.25 × 10<sup>-5 </sup> AU(empty cell, 254 nm, 1.0s)
Drift ≤ 0.5 × 10<sup>-4 </sup> AU/h (empty cell, 254 nm, 60 min)
Path length 10 mm(analytical cell)
Light Source Deuterium Lamp and Tungsten Lamp
Data-processing Workstation System contain the tables for regular samples<br/> Equipped with Internal and external standard calculation process<br/> Automatic display, print, export of data and starting of another program on completion on measurement<br/> Observe with the condition of 21 CFR part 11 directive of the FDA
System Kits 1 Set<br/> Combined A/D converters<br/> Combined vacuum degasser and 4 channel
Pump Specifications
Solvent delivery Double plunger series mode, floating plunger design, single stroke volume 30 µl
Pump Head 10 ml stainless steel pump head
Flow rate range 0.001 to 9.999 ml / min
Flow rate increment 0.001 ml/min
Maximum pressure 42 MPa
Column size 4.6 × 250 mm 5 µm
Pipeline Size 1/16 inch
Flow rate accuracy ± 0.5 %
Flow rate precision RSD ≤ 0.1%
Gradient accuracy % ± 0.5 %
Gradient repeatability % RSD ≤ 0.15%
GLP Automatic recording of pump sealing ring working time
Display LCD backlight display
Power 80 W
Power supply 220 V ± 10%, 50 Hz
Dimension 398 × 149 × 267 mm
Detector Specifications
Wavelength range 190 to 700 nm
Wavelength accuracy ± 1 nm
Wavelength repeatability 0.2 nm
Deuterium lamp 190 to 400 nm
Tungsten lamp 400 to 700 nm
Baseline noise ( Static ) Au 1 × 10<sup>-5</sup> AU/h at 254 nm with methanol/water= 80/20
Baseline drift ( Static ) Au / h 1.5 × 10<sup>-4</sup> AU/h at 254 nm with methanol/ water= 80/20
Time constants 16 grades adjustable
Display LCD backlight display
GLP Total number of lights, lamp replacement time and product serial number
Power 75 W
Power supply 220 V ± 10%, 50 Hz
Dimension (L × W × H) 398 × 149 × 267 mm
Detector Specifications
Model No HPD-A20 HPD-A21
Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm )
Wavelength accuracy ± 0.5 nm ± 2 nm
Wavelength repeatability 0.15 nm 0.3 nm
Band width 8 nm 8 nm
Optical path 10 nm 10 nm
Deuterium lamp 190 to 400 nm 190 to 400 nm
Tungsten lamp 400 to 800 nm 400 to 800 nm
Baseline noise ( Static ) Au ≤ ± 0.35 × 10-5 ≤ ± 0.5 × 10-5
Baseline drift ( Static ) Au / h ≤ 1 × 10-4 ≤ 1.5 × 10-4
Minimum detection limit ≤ 4 × 10-8 ≤ 4 × 10-8
Power 100 W 100 W
Power supply 90 V / 50 to 60 Hz 90 V / 50 to 60 Hz
Dimension 460 × 315 × 180 mm 460 × 315 × 180 mm
Weight 9.5 kg 9.5 kg
Pump Specifications
Flow rate range 0.0001 to 10 ml / min
Maximum pressure 50 MPa , 0 to 2 ml / min
Column size 4.6 × 250 mm
Sample size 5 µm
Pressure pulsation ≤ 1 % at 10 MPa
Flow rate accuracy ± 0.15 % at 1 ml / min
Flow rate repeatability % ≤ 0.1
Gradient accuracy % ± 1.5
Gradient repeatability % ≤ 0.2
Dimension 460 × 310 × 180 mm
Power supply 90 V
Weight 11.6 kg
Pump Specifications
Solvent delivery Double plunger series mode, floating plunger design, single stroke volume 30 µl
Pump Head 10 ml stainless steel pump head
Flow rate range 0.001 to 9.999 ml / min
Flow rate increment 0.001 ml/min
Maximum pressure 42 MPa
Column size 4.6 × 250 mm 5 µm
Pipeline Size 1/16 inch
Flow rate accuracy ± 0.5 %
Flow rate precision RSD ≤ 0.1%
Gradient accuracy % ± 0.5 %
Gradient repeatability % RSD ≤ 0.15%
GLP Automatic recording of pump sealing ring working time
Display LCD backlight display
Power 80 W
Power supply 220 V ± 10%, 50 Hz
Dimension 398 × 149 × 267 mm
Detector Specifications
Wavelength range 190 to 700 nm
Wavelength accuracy ± 1 nm
Wavelength repeatability 0.2 nm
Deuterium lamp 190 to 400 nm
Tungsten lamp 400 to 700 nm
Baseline noise ( Static ) Au 1 × 10<sup>-5</sup> AU/h at 254 nm with methanol/water= 80/20
Baseline drift ( Static ) Au / h 1.5 × 10<sup>-4</sup> AU/h at 254 nm with methanol/ water= 80/20
Time constants 16 grades adjustable
Display LCD backlight display
GLP Total number of lights, lamp replacement time and product serial number
Power 75 W
Power supply 220 V ± 10%, 50 Hz
Dimension (L × W × H) 398 × 149 × 267 mm
Detector Specifications
Model No HPD-A20 HPD-A21
Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm )
Wavelength accuracy ± 0.5 nm ± 2 nm
Wavelength repeatability 0.15 nm 0.3 nm
Band width 8 nm 8 nm
Optical path 10 nm 10 nm
Deuterium lamp 190 to 400 nm 190 to 400 nm
Tungsten lamp 400 to 800 nm 400 to 800 nm
Baseline noise ( Static ) Au ≤ ± 0.35 × 10-5 ≤ ± 0.5 × 10-5
Baseline drift ( Static ) Au / h ≤ 1 × 10-4 ≤ 1.5 × 10-4
Minimum detection limit ≤ 4 × 10-8 ≤ 4 × 10-8
Power 100 W 100 W
Power supply 90 V / 50 to 60 Hz 90 V / 50 to 60 Hz
Dimension 460 × 315 × 180 mm 460 × 315 × 180 mm
Weight 9.5 kg 9.5 kg
Pump Specifications
Flow rate range 0.0001 to 10 ml / min
Maximum pressure 50 MPa , 0 to 2 ml / min
Column size 4.6 × 250 mm
Sample size 5 µm
Pressure pulsation ≤ 1 % at 10 MPa
Flow rate accuracy ± 0.15 % at 1 ml / min
Flow rate repeatability % ≤ 0.1
Gradient accuracy % ± 1.5
Gradient repeatability % ≤ 0.2
Dimension 460 × 310 × 180 mm
Power supply 90 V
Weight 11.6 kg
Pump Specifications
Model No HPP-B1
Flow rate range 0.0001 to 10 ml / min
Maximum pressure 42 MPa , 0 to 5 ml / min
Column size 4.6 × 250 mm
Sample size 5 μm
Pressure pulsation ≤ 1.5 % at 10 MPa
Flow rate accuracy ± 0.2 % at 1 ml / min
Flow rate repeatability % ≤ 0.15
Gradient accuracy % ± 2
Gradient repeatability % ≤ 0.2
Dimension 450 × 260 × 190 mm
Power supply 90 V to 240 V
Weight 11.2 kg
Detector Specifications
Model No HPD-B2
Wavelength range 190 to 800 nm
Wavelength accuracy ± 0.75 nm
Wavelength repeatability 0.2 nm
Band width 8 nm
Optical path 10 nm
Deuterium lamp 190 to 400 nm
Tungsten lamp 400 to 800 nm
Baseline noise ( Static ) Au ≤ ± 0.5 × 10 <sup>-5</sup>
Baseline drift ( Static ) Au / h ≤ 1.5 x 10 <sup>-4</sup>
Minimum detection limit ≤ 4 x 10 <sup>-8</sup>
Power 100 W
Power supply 90 V / 50 to 60 Hz
Dimension 460 × 260 × 190 mm
Weight 9.2 kg
High pressure pump High pressure constant-current infusion pump
Infusion Way Micro volume series dual-piston
Infusion pressure 0 - 9999 Psi
Flow rate ≤ 0.5 %
High pressure pump Two sets of pump
Flow rate range 0.001 - 9.999 mL / min (0.001 mL / min increments)
Pressure range ≤ 0.5 MPa (0 - 42 MPa)
Flow rate precision RSD < 0.06 %
High pressure pump Single pump
Flow rate range 0.001 - 9.999 mL / min (0.001 mL / min increment)
Pressure range ≤ 0.5 MPa (0 - 42 MPa)
Flow rate precision RSD < 0.06 %