Chromatographic Pump |
- Type: High-pressure and low –pulse double plunger tandem advection pump
- Maximum Pressure: 35 MPa (PEEK)
- Flow Range: 0.001 to 9.999 mL/min
- Flow accuracy: ± 0.5 %
- Flow Repeatability: RSD ≤ 0.1%
- Flow stability: (0.2 to 0.5) mL/min, ≤ 3 %<br/>
(0.5 to 1.0) mL/min, ≤ 2 %<br/>
>1.0 mL/min, ≤ 2 %
|
Built-in Heat conduction type column heater |
- Temperature range: Room temperature +5°C to 60°C (41 to 140°F)
- Allowable deviation: ± 1°C
- Temperature stability: ≤ 0.5°C/h
|
Electromagnetic injection valve |
- Maximum pressure: 35 MPa
- Control mode: By stepper motor
- Power supply: 24 V (DC)
|
Digital and temperature control detection system |
- Type: Temperature control and bipolar conductivity detector
- Detection type: Bipolar conductivity detection
- Cell volume: ≤ 0.8 µL
- Measurement range: 0 to 45000 µ S/cm (adjustable)
- Detection resolution: ≤ 0.0020 nS/cm
- Output voltage: -6000 to 6000 mV (adjustable)
- Temperature range: Room temperature +5°C to 60°C (41 to 140°F)
- Temperature accuracy: ± 0.01°C
- Maximum pressure: 10 MPa
- Baseline drift: ≤ 20 % FS/30 min
- Baseline noise: ≤0.5 % FS
- Instrument linearity: ≥ 0.999
- Quantitative Repeatability: ≤0.5 %
- Qualitative Repeatability: ≤2 %
- Minimum detectable concentration: Cl-≤0.005 µg/mL<br/>
Li+ ≤0.005 µg/mL
|
Flow system |
- Plastic path flow: Made of PEEK material
- Six way valve: PEEK material pressure 5000 psi, independent automatic collecting and flow function
|
Panel computer |
- Display screen: 12.3 inch
- Internal memory: 2G
- Weight: 786 g
|
External size (L × W × H) |
440 × 226 × 420 mm |
Net weight |
8 kg |
Gross weight |
11 kg |
- Built-in column temperature control box for accurate experimental results
- Capable with Flash column for rapid detection of anions and cations
- Intelligent flow path clearing system for easy cleaning
- Real time display
- Wi-Fi connection for convenient data transfer
- Continuous, non-stop 10 hours working system
- Easy to operate
Accessories No. |
Name |
Qty. |
1 |
Anion Column |
1 |
2 |
Guard Column |
1 |
3 |
Anion Suppressor |
1 |
4 |
Spare Parts Kit |
1 |
Used for on field selective analysis of organic and inorganic ions, also used in quality and research laboratories across pharmaceutical, water quality, chemical industries.