| Chamber volume |
240 L |
| Heating method |
Water Jacketed |
| CO<sub>2</sub> control range |
0 to 20% |
| CO<sub>2</sub> control accuracy |
±0.1% (IR sensor) |
| CO<sub>2</sub> restore time |
≤3min (Door open 30s, recovery to 5%) |
| Temperature control range |
RT +5°C to 50°C |
| Ambient temperature |
+5° to 30°C |
| Temperature accuracy |
±0.1°C |
| Temperature restore time |
≤8min (Door open 30s, recovery to 37°C) |
| Related humidity |
Nature vaporate > 95% (Can equip with relative humidity digital display) |
| No. of shelves |
3 |
| Sterilization |
UV sterilization; HEPA sterilization |
| Fan |
Cycle fan with automatic speed and control |
| Controller |
Touch screen |
| Curve |
On time performance curve |
| Screen lock |
Automatic |
| CO<sub>2</sub> concentration sensor |
IR sensor |
| Temperature sensor |
PT100 |
| Door heating system |
Yes |
| Over temperature protection system |
Yes |
| CO<sub>2</sub> inlet control system |
Yes |
| Alarm system |
High/low and over-temperature; door temperature sensor failure; over-temperature sensor failure; CO2 condensation too high or too low; independent temperature limiter, Door open too long and Disinfection and sterilization status reminder |
| Communication port |
RS485 |
| Chamber dimensions (W×D×H) |
600×520×780 mm |
| External dimensions (W×D×H) |
754×760×1020 mm |
| Packing dimension (W×D×H) |
910×940×1230 mm |
| Gross weight |
148 kg |
| Input power |
1000 W |
| Power supply |
AC 220 V/50 Hz |
- Intelligent touch screen controller display time performance curve for temperature, humidity (option) and CO2 concentration
- Door open/lock message; abnormal alarm; Automatic screen lock when parameters are set
- 72 hours machine inquiry for user convenience with RS-485 communication port
- CO2 concentration (IR) sensor with 0.1% accuracy; PT100 temperature sensor
- Over temperature protection system; Outer door ring with heating functions
- UV and HEPA sterilization
- Micro biological HEPA filter at CO2 access port, can filters particles of diameter ≥0.3um
- Cycle fan with automatic speed and control, avoids sample evaporation
- CO2 inlet control system with pressure release valve and protection system
- Alarm system: High/low and over-temperature; door temperature sensor failure; over-temperature sensor failure; CO2 condensation too high or too low; independent temperature limiter, Door open too long and Disinfection and sterilization status reminder
- Stable, reliable and accurate, beneficial with multiple door openings or power outages
Humidity display
CO2 pressure valve
RS485
Water jacketed CO2 Incubator are used for tissue engineering, cell culture, in vitro fertilization, diagnostics, neuroscience, cancer research and for developing biosensors across medical research and the pharmaceutical industry.