| Temperature Control Range |
0°C to 100°C if Ambient temp. ≤ 20°C<br/>
4°C to 100°C if Ambient temp. ≤ 25°C<br/>
10°C to 100°C if Ambient temp .≤ 30°C |
| Temperature Setting Range |
0°C to 100°C |
| Timer |
1 min to 99 h 59 min |
| Temperature Control Accuracy |
± 0.5°C |
| Display Accuracy |
± 0.1°C |
| Mixing Speed |
200 to 1500 rpm |
| Heating Time (25 to 100°C) |
≤ 15 min |
| Cooling Time |
≤ 30 min (from RT. to RT.-20°C)<br/> ≤ 15 min (from 100°C to 20°C) |
| Mixing Orbit |
2 mm |
| Dimension |
300 × 220 × 175 mm |
| Net Weight |
8.5 kg |
- Performs heating, mixing and cooling of samples
- Peltier device for better thermal control
- Standard platform accommodates various sample block
- LCD Display
- Accurate control of speed, temperature and time
- Over-heat protection
- Least noise generated
- Easy to clean design
- Meets CE standards
- Compact Structure
Various preparatory steps in PCR, ELISA, genetic analysis, biochemical assays, cell lysis, cell biology studies, DNA/RNA isolation and purification, re-suspension of pellet etc. require it.