| Monochromator |
|
| Wavelength range |
180 - 900 nm |
| Spectral Bandwidth |
0, 0.2, 0.4, 1.0, 2.0nm (Automatic setting) |
| Wavelength accuracy |
≤ 0.15 nm |
| Wavelength repeatability |
± 0.1 nm |
| Baseline drift |
≤ ± 0.002 Abs / 30 min (Static), ≤ ± 0.005 Abs / 30 min (Dynamic) |
| Grating groove density |
|
| Sample Volume |
|
| The volume of sample |
|
| Repeated Sampling Frequency |
|
| Cleaning and waste container volume |
|
| Absorbance Range |
0 - 4 A |
| The Program |
|
| Temperature Range |
|
| Detector |
Imported photomultiplier tube |
| Heating Mode |
|
| Temperature Control Method |
|
| Temperature Control |
|
| Resolution |
|
| Lamp holder |
≤ 6 lamps automatic turret, automatic alignment |
| Lamp current adjustment range |
|
| Lamp power supply mode |
pulsed power supply |
| Optical System |
large 1800 /mm grating ruling, full closed optical system |
| Atomization Chamber |
Polymer explosion-proof spray chamber |
| Characteristics Volume |
|
| Characteristic concentration (Cu) |
0.015 μg / mL / 1 % |
| Detection Limit |
|
| Detection Limit (Cu) |
0.002 μg/mL |
| Background calibration ability |
|
| Burner |
Full titanium combustion head, 50 mm or 100 mm general combustion head |
| Nebulizer |
Atomizer efficient glass atomizer,can also be customized |
| Precision |
RSD ≤ 0.5 % |
| Ignition Type |
Microcomputer control, automatic ignition |
| Sprayer |
|
| Room Spray |
|
| Sample Tray |
|
| The Smallest Lincrement |
|
| Safety measures |
|
| Gas Control |
Automatic gas control system |
| Measurement |
Flame method, Hydride generation-Atomic absorption method |
| Concentration calculation |
Standard curve method(1 - 3 times curve), automatic fitting |
| The Cooling Water |
|
| The number of repeated measurements |
1 to 99, automatic calculation of average concentration, standard deviation, relative standard deviation. |
| Print results |
Print the final stages of test data or analysis. The results can be exported into Excel worksheet. |
| Software Environment |
|
| Operating temperature |
|
| Working humidity |
|
| Power supply |
110 / 220 V (+5 % ~ (-10) %), 60 / 50 Hz; 5000 VA |
| Power (VAC) |
|
| Gross Weight |
|
| Weight |
|
| Dimension (L × W × H) |
|
- 1800 /mm reticle (dispersion rate) large area grating Monochromator
- Titanium burner for the flame analysis
- Polymer automization chamber
- Automatic analysis ensures sensitivity and reproducibility of the flame method
- Light source 6 lamp holder
- Graphic furnace system gives information on analysis of the whole process
- Convenient data processing & sharing & printing
- Powerful software with automatic instrument and additional control
Atomic absorption spectrophotometer is widely used in analysis of elements from the mixture in scientific research, environmental science, metallurgy, agriculture & chemical industry, food, mining, and petrochemical, pharmaceutical industries.