Benchtop Ion meter for Cu, LBIM-A27

Benchtop Ion meter for Cu, LBIM-A27

Benchtop Ion meter for Cu, LBIM-A27 comes with (Cu) cupric ion selective electrode for cupric ion concentration measurement. This Ion meter comes with ± 1 % F.S. measurement accuracy. Its direct ion concentration readout helps in elaborate measurement process. The system menu allows setting the 8 parameters like calibration points, stability condition etc. It can measure concentration in ppm, mg / L, mol / L. The mV measurement is used to check the performance of ion selective electrode.

Specifications

Features

Applications

Download Catalog

Ion measurement rangeCu range: 1×10<sup>8</sup> - 1M, 0.06-6400 ppm
Ion measurement accuracy± 1 % F.S.
Resolution0.001, 0.01, 0.1, 1
Calibration points2 to 5 points
Calibration solutions0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000 ppm, mol / L, mg / L
mV range(-1999.9) to 1999.9 mV
mV accuracy± 0.2 mV
mV resolution0.1 mV
Temperature measurement range0 to 105 °C, 32 to 221 °F
Temperature accuracy± 0.5 °C, ± 0.9 °F
Temperature resolution0.1°C
Temperature compensation0 to 100 °C, 32 to 212 °F
Compensation modeManual or automatic
Data hold functionManual or automatic endpoint detection
Reset functionYes
Stability conditionLow or high
Calibration due0 to 31 days
Power offManual or automatic (10,20,30 minutes)
MemoryStores up to 500 sets of data
OutputUSB communication interface
ConnectorBNC
DisplayLCD (130 × 110 mm)
PowerDC 5 V, using AC adapter, AC 220 V / 50 Hz
Dimension (L × W × H)210 × 188 × 60 mm
Weight1.5 kg

Related Products of Benchtop Ion Meter

Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.

Benchtop Ion meter for Ca, LBIM-A13

Benchtop Ion meter for Ca, LBIM-A13

Ion measurement rangeCa range: 5 × 10<sup>7</sup> - 1 M, 0.02 - 4000 ppm
Ion measurement accuracy± 0.5 % F.S.
Resolution0.001, 0.01, 0.1, 1
Benchtop Ion meter for Ag, LBIM-A22

Benchtop Ion meter for Ag, LBIM-A22

Ion measurement rangeAg range: 1 × 10<sup>7</sup> - 1 M, 0.01 - 107900 ppm
Ion measurement accuracy± 1 % F.S.
Resolution0.001, 0.01, 0.1, 1
Benchtop Ion meter for Cn, LBIM-A21

Benchtop Ion meter for Cn, LBIM-A21

Ion measurement rangeCn range: 5 × 10<sup>6</sup> – 0.01 M, 0.2 – 260 ppm
Ion measurement accuracy± 1 % F.S.
Resolution0.001, 0.01, 0.1, 1
Benchtop Ion meter for Na, LBIM-A12

Benchtop Ion meter for Na, LBIM-A12

Ion measurement rangeNa range: 1 × 10<sup>5</sup> - 1 M, 0.1 - 23000 ppm
Ion measurement accuracy± 0.5 % F.S.
Resolution0.001, 0.01, 0.1, 1
Benchtop Ion meter for NH3 , LBIM-A20

Benchtop Ion meter for NH3 , LBIM-A20

Ion measurement rangeNH<sub>3</sub> range: 1 × 10<sup>6</sup> - 1 M, 0.02 - 17000 ppm
Ion measurement accuracy± 1 % F.S.
Resolution0.001, 0.01, 0.1, 1
Benchtop Ion meter for Pb, LBIM-A25

Benchtop Ion meter for Pb, LBIM-A25

Ion measurement rangePb range: 1×10<sup>8</sup> – 0.1M, 0.2-20700 ppm
Ion measurement accuracy± 1 % F.S.
Resolution0.001, 0.01, 0.1, 1
Benchtop Ion meter for F, LBIM-A11

Benchtop Ion meter for F, LBIM-A11

Ion measurement rangeF range: 1 × 10<sup>6</sup>M – saturation, 0.02 ppm to saturation
Ion measurement accuracy± 0.5 % F.S.
Resolution0.001, 0.01, 0.1, 1
Benchtop Ion  meter  for K, LBIM-A23

Benchtop Ion meter for K, LBIM-A23

Ion measurement rangeK range: 1×10<sup>6</sup> - 1M, 0.04 -39000 ppm
Ion measurement accuracy± 1 % F.S.
Resolution0.001, 0.01, 0.1, 1
Benchtop Ion meter for NH4, LBIM-A15

Benchtop Ion meter for NH4, LBIM-A15

Ion measurement rangeNH<sub>4</sub> range: 5 × 10<sup>6</sup> – 1 M, 0.1 – 18000 ppm
Ion measurement accuracy± 0.5 % F.S.
Resolution0.001, 0.01, 0.1, 1
Benchtop Ion meter for I, LBIM-A26

Benchtop Ion meter for I, LBIM-A26

Ion measurement rangeI range: 1×10<sup>8</sup>- 1M, 0.06 -127000 ppm
Ion measurement accuracy± 1 % F.S.
Resolution0.001, 0.01, 0.1, 1
Benchtop Ion meter for Cl, LBIM-A10

Benchtop Ion meter for Cl, LBIM-A10

Ion measurement rangeCl range: 5 × 10<sup>6</sup> - 1 M, 1.8 - 35500 ppm
Ion measurement accuracy± 0.5 % F.S.
Resolution0.001, 0.01, 0.1, 1
Benchtop Ion meter for S, LBIM-A24

Benchtop Ion meter for S, LBIM-A24

Ion measurement rangeS range: 1×10<sup>7</sup> - 1M, 0.003-32100 ppm
Ion measurement accuracy± 1 % F.S.
Resolution0.001, 0.01, 0.1, 1
Benchtop Ion meter for Br, LBIM-A14

Benchtop Ion meter for Br, LBIM-A14

Ion measurement rangeBr range: 5 × 10<sup>6</sup> - 1 M, 0.4 - 79900 ppm
Ion measurement accuracy± 0.5 % F.S.
Resolution0.001, 0.01, 0.1, 1
Benchtop Ion meter for Cd, LBIM-A28

Benchtop Ion meter for Cd, LBIM-A28

Ion measurement rangeCd range: 1×10<sup>6</sup>– 1M, 0.01-11200 ppm
Ion measurement accuracy± 1 % F.S.
Resolution0.001, 0.01, 0.1, 1

pH Meter (Benchtop)

Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.

Benchtop pH meter LBPM-A20

Benchtop pH meter LBPM-A20

pH measurement range-1.00 to 15.00 pH
pH accuracy± 0.01 pH
pH resolution0.01 pH
Benchtop pH meter LBPM-B10

Benchtop pH meter LBPM-B10

pH measurement range-2.000 to 20.000 pH
pH accuracy± 0.002 pH
Benchtop pH Meter LBPM-B11

Benchtop pH Meter LBPM-B11

Temperature compensation0 to 100°C, 32°F to 212°F Manual or Automatic
Hold functionManual or automatic endpoint detection
Stability criteriaStandard or High-accuracy
Benchtop pH meter LBPM-A11

Benchtop pH meter LBPM-A11

pH measurement range0.00 to 14.00 pH
pH accuracy± 0.01 pH
pH resolution0.01 pH

Benchtop Conductivity Meter

Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.

Benchtop Conductivity Meter LBCM-A13

Benchtop Conductivity Meter LBCM-A13

Conductivity measurement range0.000 μS/cm to 3000mS/cm
Resistivity measurement range5.00 Ω·cm to 100.0 MΩ·cm
TDS measurement range0.000 mg/L to 1000g/L
Benchtop conductivity meter LBCM-A11

Benchtop conductivity meter LBCM-A11

Conductivity measurement range0 - 20.0, 200.0, 2000 µS / cm, 200.0 mS / cm
Conductivity accuracy± 1 % F.S.
Conductivity resolution0.01, 0.1, 1
Benchtop Conductivity Meter LBCM-A14

Benchtop Conductivity Meter LBCM-A14

Conductivity Mode
Conductivity measurement range: 0.01-20.0, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
Conductivity accuracy: ±1% F.S.
Conductivity resolution: 0.001, 0.01, 0.1, 1
Conductivity calibration points: 1 to 3 points
Conductivity calibration solutions: 10 µS/cm , 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm
TDS Mode
TDS measurement Range: 0 to 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
TDS Accuracy: ± 1 % F.S.
TDS Factor: 0.01 to 1.00 (default 0.5)
TDS Calibration points: 3 points
Temperature Mode
Temperature range: 0 to 105°C, 32 to 221°F
Temperature accuracy: ±1°C
Temperature resolution: 0.1°C, 0.1°F
Temperature calibration points: 1 point
Temperature Compensation Range and Mode: Manual or Automatic, 0 to 100°C , 32 to 212°F
Temperature coefficient: 0.0 to 10.0 % per °C

Benchtop Dissolved Oxygen Meter

Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.

Benchtop Dissolved oxygen meter LBDO-A11

Benchtop Dissolved oxygen meter LBDO-A11

Dissolved oxygen range0.0 to 20.0 mg / L (or ppm)
Dissolved oxygen accuracy± 0.2 mg / L
Resolution0.01 mg / L
Benchtop Dissolved oxygen meter LBDO-A10

Benchtop Dissolved oxygen meter LBDO-A10

Dissolved oxygen range0.0 to 20.0 mg / L (or ppm)
Dissolved oxygen accuracy± 0.5 mg / L
Resolution0.1 mg / L

Benchtop Water Hardness Meter

Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.

Benchtop Water hardness meter LBWM-A10

Benchtop Water hardness meter LBWM-A10

Water hardness range0.05 mmol / L -0.2 mol / L <br/> (CaCO<sub>3</sub>) 0 - 19999 mg / L, (CaO) 0 - 11220 mg / L, (boiler) 0 - 400 mmol / L, (Ca) 0 - 8020 mg / L<br/> (German degree) 0 - 1122 °dH, (English degree) 0 - 1404 °e, (French degree) 0 - 2000 °f
Accuracy± 1 % F.S.
Calibration points2 to 5 points

Conductivity Tester

Labtron pocket testers are small and compact in size, gives fast and accurate results. The Pocket testers consist of pH, ORP, TDS, dissolved oxygen, salinity and conductivity tester for various laboratory as well as outdoor applications. Measurements can be taken by inserting the probe in to the test solution and after gentle stirring, stable readings can be noted down.

Pocket Conductivity tester LPCT-A11

Pocket Conductivity tester LPCT-A11

Conductivity measurement range10 to 1999 µS/cm
Type of conductivityMid-range conductivity
Conductivity accuracy± 1 % F.S.
Pocket Conductivity tester LPCT-A12

Pocket Conductivity tester LPCT-A12

Conductivity measurement range0.1 to 19.99 mS/cm
Type of conductivityHigh range conductivity
Conductivity accuracy± 1 % F.S.
Pocket Conductivity tester LPCT-A10

Pocket Conductivity tester LPCT-A10

Conductivity measurement range1.0 to 199.9 µS/cm
Type of conductivityLow range conductivity
Conductivity accuracy± 1 % F.S.

Dissolved Oxygen Tester

Labtron pocket testers are small and compact in size, gives fast and accurate results. The Pocket testers consist of pH, ORP, TDS, dissolved oxygen, salinity and conductivity tester for various laboratory as well as outdoor applications. Measurements can be taken by inserting the probe in to the test solution and after gentle stirring, stable readings can be noted down.

Pocket Dissolved oxygen tester LPKDO-A10

Pocket Dissolved oxygen tester LPKDO-A10

Dissolved oxygen range0.0 to 20.0 mg/L (or ppm)
Dissolved oxygen accuracy± 0.5 mg/L
% of saturation of oxygen0.0 to 200.0%

ORP Tester

Labtron pocket testers are small and compact in size, gives fast and accurate results. The Pocket testers consist of pH, ORP, TDS, dissolved oxygen, salinity and conductivity tester for various laboratory as well as outdoor applications. Measurements can be taken by inserting the probe in to the test solution and after gentle stirring, stable readings can be noted down.

Pocket ORP tester LPOT-A11

Pocket ORP tester LPOT-A11

ConnectorBNC
ORP measurement range(-999) to 999 mV
Pocket ORP tester LPOT-A10

Pocket ORP tester LPOT-A10

Sensor typePlatinum sensor
Connector6 pin
ORP measurement range(-999) to 999 mV

pH Tester

Labtron pocket testers are small and compact in size, gives fast and accurate results. The Pocket testers consist of pH, ORP, TDS, dissolved oxygen, salinity and conductivity tester for various laboratory as well as outdoor applications. Measurements can be taken by inserting the probe in to the test solution and after gentle stirring, stable readings can be noted down.

Pocket pH tester LPPT-A32

Pocket pH tester LPPT-A32

pH measurement rangepH range: (-1.00) to 15.00 pH
Shape of electrodeLong distance planar pH electrode
pH measurement accuracy± 0.01 pH
Pocket pH tester LPPT-A30

Pocket pH tester LPPT-A30

pH measurement rangepH range: (-1.00) to 15.00 pH
Shape of electrodeRound glass pH electrode
pH measurement accuracy± 0.01 pH
Pocket pH tester LPPT-A10

Pocket pH tester LPPT-A10

pH measurement range0.0 - 14.00 pH
Shape of electrodeRound glass pH electrode
pH measurement accuracy± 0.1 pH
Multi purpose Pocket pH tester LPPT-B10

Multi purpose Pocket pH tester LPPT-B10

pH measurement rangepH range: (-1.00) to 15.00 pH
Shape of electrodeDifferent electrodes can be connceted via BNC Connector
pH measurement accuracy± 0.01 pH
Pocket pH tester LPPT-A20

Pocket pH tester LPPT-A20

pH measurement rangepH range: (-1.00) to 15.00 pH
Shape of electrodeRound glass pH electrode
pH measurement accuracy± 0.01 pH
Pocket pH tester LPPT-A12

Pocket pH tester LPPT-A12

pH measurement range0.0 - 14.00 pH
Shape of electrodeLong distance planar pH electrode
pH measurement accuracy± 0.1 pH
Pocket pH tester LPPT-A11

Pocket pH tester LPPT-A11

pH measurement range0.0 - 14.00 pH
Shape of electrodeFlat pH electrode
pH measurement accuracy± 0.1 pH
Pocket pH tester LPPT-A21

Pocket pH tester LPPT-A21

pH measurement rangepH range: (-1.00) to 15.00 pH
Shape of electrodeFlat pH electrode
pH measurement accuracy± 0.01 pH
Pocket pH tester LPPT-A31

Pocket pH tester LPPT-A31

pH measurement rangepH range: (-1.00) to 15.00 pH
Shape of electrodeFlat pH electrode
pH measurement accuracy± 0.01 pH
Pocket pH tester LPPT-A22

Pocket pH tester LPPT-A22

pH measurement rangepH range: (-1.00) to 15.00 pH
Shape of electrodeLong distance planar pH electrode
pH measurement accuracy± 0.01 pH

Salinity Tester

Labtron pocket testers are small and compact in size, gives fast and accurate results. The Pocket testers consist of pH, ORP, TDS, dissolved oxygen, salinity and conductivity tester for various laboratory as well as outdoor applications. Measurements can be taken by inserting the probe in to the test solution and after gentle stirring, stable readings can be noted down.

Pocket Salinity tester LPST-A10

Pocket Salinity tester LPST-A10

Salinity measurement range0.00 to 10.00 ppt
Salinity accuracy± 1 % F.S
Conductivity range0 - 20.00, 200.0, 2000 µS/cm, 20.00 mS/cm
Pocket Salinity tester LPST-A11

Pocket Salinity tester LPST-A11

Salinity measurement range0.00 to 80.00 ppt
Salinity accuracy± 1 % F.S
Conductivity range0 - 200.0 µS/cm, 100 - 2000 µS/cm, 20.00 mS/cm

TDS Tester

Labtron pocket testers are small and compact in size, gives fast and accurate results. The Pocket testers consist of pH, ORP, TDS, dissolved oxygen, salinity and conductivity tester for various laboratory as well as outdoor applications. Measurements can be taken by inserting the probe in to the test solution and after gentle stirring, stable readings can be noted down.

Multi-range pocket TDS tester LPTS-A20

Multi-range pocket TDS tester LPTS-A20

TDS range0 - 20 ppt
TDS accuracy± 1 % F.S.
TDS factor0.1 to 1.0 (default 0.5)
Pocket TDS tester LPTS-A11

Pocket TDS tester LPTS-A11

TDS range5 - 1000 ppm
TDS accuracy± 1 % F.S.
TDS factor0.4 to 1.0 (default 0.5)
Pocket TDS tester LPTS-A10

Pocket TDS tester LPTS-A10

TDS range0.5 - 100.0 ppm
TDS accuracy± 1 % F.S.
TDS factor0.4 to 1.0 (default 0.5)
Pocket TDS tester LPTS-A12

Pocket TDS tester LPTS-A12

TDS range0.05 - 10.00 ppt
TDS accuracy± 1 % F.S.
TDS factor0.4 to 1.0 (default 0.5)

Portable Dissolve Oxygen Meter

Labtron portable meters are used in quality analysis of water and other solutions. With portability, waterproof features, finds applications in indoor as well as outdoor measurement of pH, ORP, dissolved oxygen, water hardness, salinity, TDS etc.

Portable dissolved oxygen meter LPRDO-A11

Portable dissolved oxygen meter LPRDO-A11

Dissolved oxygen range0.0 - 20.0 mg / L (or ppm)
Dissolved oxygen accuracy± 0.2 mg / L
Resolution0.01 mg / L
Portable Dissolved oxygen meter LPRDO-A10

Portable Dissolved oxygen meter LPRDO-A10

Dissolved oxygen range0.0 to 20.0 mg / L (or ppm)
Dissolved oxygen accuracy± 0.5 mg / L
Resolution0.1 mg / L

Portable Multi Parameter Water Quality Meter

Labtron portable meters are used in quality analysis of water and other solutions. With portability, waterproof features, finds applications in indoor as well as outdoor measurement of pH, ORP, dissolved oxygen, water hardness, salinity, TDS etc.

Portable Multi-parameter Water Quality Meter LMPWM-A21

Portable Multi-parameter Water Quality Meter LMPWM-A21

pH mode
pH measurement range: (-2.000) to 20.000 pH
pH accuracy: ± 0.002 pH
pH calibration points: 1 - 5 points
pH calibration solutions: USA, NIST, DIN or user defined
ORP measurement
ORP measurement range: -1999.9 to 1999.9 mV
ORP accuracy: ±0.2 mV
ORP calibration points: 1 point
ORP calibration solutions: mV & relative mV
Dissolved oxygen
DO concentration range: 0.00 to 20.00 mg/L
DO accuracy: ± 0.2 mg/L
Calibration points: 1 - 2 points
Pressure correction: 60.0 to 112.5 kPa, 450 - 850 mmHg
Salinity correction: 0 to 50 g/L
% Saturation of Oxygen: 0.0 to 200.0 %
Accuracy: ± 2.0 %
Portable Multi-parameter Water Quality Meter LMPWM-A11

Portable Multi-parameter Water Quality Meter LMPWM-A11

pH mode
pH measurement range: (-2.000) to 20.000 pH
pH accuracy: ± 0.002 pH
pH calibration points: 1 - 5 points
pH calibration solutions: USA, NIST, DIN or user defined
ORP measurement
ORP measurement range: -1999.9 to 1999.9 mV
ORP accuracy: ±0.2 mV
ORP calibration points: 1 point
ORP calibration solutions: mV & relative mV
Ion measurement
Ion measurement range:
Ion accuracy:
Ion calibration points:
Ion calibration solutions:
Portable Multi-parameter Water Quality Meter LMPWM-A20

Portable Multi-parameter Water Quality Meter LMPWM-A20

pH mode
pH measurement range: (-2.000) to 20.000 pH
pH accuracy: ± 0.002 pH
pH calibration points: 1 - 5 points
pH calibration solutions: USA, NIST, DIN or user defined
ORP measurement
ORP measurement range: -1999.9 to 1999.9 mV
ORP accuracy: ±0.2 mV
ORP calibration points: 1 point
ORP calibration solutions: mV & relative mV
Conductivity measurement
Conductivity measurement range: 0 - 20.00, 200.0, 2000 µS / cm, 20.00, 200.2 mS / cm
Conductivity accuracy: ± 0.5 % F.S.
Conductivity calibration points: 1 - 5 points
Conductivity calibration solutions: 10 µS / cm, 84 µS / cm, 1413 µS / cm, 12.88mS / cm, 111.8mS / cm
Temperature coefficient: 0.0 to 10.0 % / °C
Compensation modes: Linear or pre water
Cell constant: K = 0.1, 1, 10 or user-defined
Normalization temperature: 20 °C or 25 °C
Portable Multi-parameter Water Quality Meter LMPWM-A10

Portable Multi-parameter Water Quality Meter LMPWM-A10

pH mode
pH measurement range: (-2.000) to 20.000 pH
pH accuracy: ± 0.002 pH
pH calibration points: 1 - 5 points
pH calibration solutions: USA, NIST, DIN or user defined
ORP measurement
ORP measurement range: -1999.9 to 1999.9 mV
ORP accuracy: ±0.2 mV
ORP calibration points: 1 point
ORP calibration solutions: mV & relative mV
Ion measurement
Ion measurement range: 0 to 19999 ppm, mg/L, mol/L
Ion accuracy: ± 0.5 % F.S. (monovalent), ± 1 % F.S. (divalent)
Ion calibration points: 2 - 5 points
Ion calibration solutions: 0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000 ppm, mol/L, mg/L

pH Meters (Portable)

Labtron portable meters are used in quality analysis of water and other solutions. With portability, waterproof features, finds applications in indoor as well as outdoor measurement of pH, ORP, dissolved oxygen, water hardness, salinity, TDS etc.

Portable pH meter LPRPM-A10

Portable pH meter LPRPM-A10

pH range(-2.00) to 16.00 pH
pH accuracy± 0.01 pH
pH resolution0.01 pH
Portable pH/ORP meter LPRPM-A11

Portable pH/ORP meter LPRPM-A11

pH range(-2.00) to 20.00 pH
pH accuracy± 0.002 pH
pH resolution0.1, 0.01, 0.001 pH

Portable Water Hardness Meters

Labtron portable meters are used in quality analysis of water and other solutions. With portability, waterproof features, finds applications in indoor as well as outdoor measurement of pH, ORP, dissolved oxygen, water hardness, salinity, TDS etc.

Portable water hardness meter LPWM-A10

Portable water hardness meter LPWM-A10

Water hardness range0.05 mmol / L - 0.2 mol / L<br/> (CaCO<sub>3</sub>) 0 - 19999 mg / L, (CaO) 0 - 11220 mg / L, (boiler) 0 - 400 mmol / L, (Ca) 0 - 8020 mg / L<br/> (German degree) 0 - 1122 °dH, (English degree) 0 - 1404 °e, (French degree) 0 - 2000 °f
Accuracy± 1 % F.S.
Calibration points2 to 5 points

Multi Parameter Water Quality Analyzer

Multi parameter water quality analyzer LMWA-A11

Multi parameter water quality analyzer LMWA-A11

COD
COD determination range: 5 to 5000 mg/ L
COD measurement accuracy: COD < 50 mg/ L, ≤ ± 10 % and COD > 50 mg/ L, ≤ ± 5 %
COD minimum detection limit: 0.1 mg/ L
COD measurement time: 20 minutes
COD repeatability: ≤ ± 5 %
Processing sample number per batch: 12 pcs
Consumables for colorimetric COD measurement: Colorimetric tubes and Cuvette
COD reactor digestion temperature: 165 ± 0.5°C
COD reactor digestion time: 10 min
COD reactor time setting switches: 3 pcs
Anti-chlorine interference: (Cl) < 1000 mg/ L No effect
Ammonia nitrogen
Ammonia nitrogen determination range: 0.02 to 30 mg/ L
Ammonia nitrogen measurement accuracy: ≤ ± 10 %
Ammonia nitrogen minimum detection limit: 0.01 mg/ L
Ammonia nitrogen measurement time: 10 to 15 minutes
Ammonia nitrogen repeatability: ≤ ± 5 %
Processing sample number per batch: 20 pcs
Consumables for colorimetric ammonia nitrogen measurement: Colorimetric tubes and Cuvette
Ammonia nitrogen reactor time setting switches: 3 pcs
Total phosphorous
Total phosphorous determination range: 0.002 to 10 mg/ L
Total phosphorous total phosphorous measurement accuracy: ≤ ± 5 %
Total phosphorous minimum detection limit: 0.001 mg/ L
Total phosphorous measurement time: 35 to 50 minutes
Total phosphorous repeatability: ≤ ± 5 %
Processing sample number per batch: 12 pcs
Consumables for colorimetric total phosphorous measurement: Colorimetric tubes and Cuvette
Total phosphorous reactor digestion temperature: 120 ± 0.5°C
Total phosphorous reactor digestion time: 30 min
Total phosphorous reactor time setting switches: 3 pcs
Multi parameter water quality analyzer LMWA-A10

Multi parameter water quality analyzer LMWA-A10

COD
COD determination range: 2 to 10000 mg/ L
COD measurement accuracy: COD < 50 mg/ L, ≤ ± 10 % and COD > 50 mg/ L, ≤ ± 5 %
COD minimum detection limit: 0.1 mg/ L
COD measurement time: 20 minutes
COD repeatability: ≤ ± 5 %
Processing sample number per batch:
Consumables for colorimetric COD measurement: Colorimetric tubes and Cuvette
COD reactor digestion temperature: 165 ± 0.5°C
COD reactor digestion time: 10 min
COD reactor time setting switches: 3 pcs
Anti-chlorine interference: (Cl) < 1000 mg/ L No effect
Ammonia nitrogen
Ammonia nitrogen determination range: 0.02 to 60 mg/ L
Ammonia nitrogen measurement accuracy: ≤ ± 10 %
Ammonia nitrogen minimum detection limit: 0.01 mg/ L
Ammonia nitrogen measurement time: 10 to 15 minutes
Ammonia nitrogen repeatability: ≤ ± 5 %
Processing sample number per batch:
Consumables for colorimetric ammonia nitrogen measurement: Colorimetric tubes and Cuvette
Ammonia nitrogen reactor time setting switches: 3 pcs
Total phosphorous
Total phosphorous determination range: 0.002 to 20 mg/ L
Total phosphorous total phosphorous measurement accuracy: ≤ ± 5 %
Total phosphorous minimum detection limit: 0.001 mg/ L
Total phosphorous measurement time: 35 to 50 minutes
Total phosphorous repeatability: ≤ ± 5 %
Processing sample number per batch:
Consumables for colorimetric total phosphorous measurement: Colorimetric tubes and Cuvette
Total phosphorous reactor digestion temperature: 120 ± 0.5°C
Total phosphorous reactor digestion time: 30 min
Total phosphorous reactor time setting switches: 3 pcs

Turbidity Meter

Portable Turbidity Meter LTM-A10

Portable Turbidity Meter LTM-A10

Measuring range (NTU)0 to 200
Minimum readout value (NTU)0.1
Calibration pointNil
Bench top Turbidity Meter LTM-B14

Bench top Turbidity Meter LTM-B14

Measuring range (NTU)0 to 50, 0 to 100, 0 to 200, 0 to 400, 0 to 800, 0 to 1000
Minimum readout value (NTU)0.01, 0.1
Calibration point6
Bench top Turbidity Meter LTM-B13

Bench top Turbidity Meter LTM-B13

Measuring range (NTU)0 to 50, 0 to 100, 0 to 200, 0 to 400, 0 to 800
Minimum readout value (NTU)0.01, 0.1
Calibration point5
Portable turbidity meter LTM-A16

Portable turbidity meter LTM-A16

Light sourceLED/860 nm (Wavelength)
Readings displayLCD
Unit switchesNTU, FTU, EBC, ASBC
Bench top Turbidity Meter (with built-in printer) LTM-C14

Bench top Turbidity Meter (with built-in printer) LTM-C14

Measuring range (NTU)0 to 10, 0 to 100, 0 to 200, 0 to 500, 0 to 1000
Minimum readout value (NTU)0.001
Calibration point5
Bench top Turbidity Meter (with built-in printer) LTM-C11

Bench top Turbidity Meter (with built-in printer) LTM-C11

Measuring range (NTU)0 to 100, 0 to 200, 0 to 500
Minimum readout value (NTU)0.01
Calibration point3
Bench top Turbidity Meter (with built-in printer) LTM-C15

Bench top Turbidity Meter (with built-in printer) LTM-C15

Measuring range (NTU)0 to 10, 0 to 100
Minimum readout value (NTU)0.001
Calibration point2
Portable Turbidity Meter LTM-A12

Portable Turbidity Meter LTM-A12

Measuring range (NTU)0 to 100, 100 to 400
Minimum readout value (NTU)0.01, 0.1
Calibration point5
Bench top Turbidity Meter (with built-in printer) LTM-C12

Bench top Turbidity Meter (with built-in printer) LTM-C12

Measuring range (NTU)0 to 10, 0 to 100, 0 to 200, 0 to 500
Minimum readout value (NTU)0.001
Calibration point4
Bench top Turbidity Meter (with built-in printer) LTM-C13

Bench top Turbidity Meter (with built-in printer) LTM-C13

Measuring range (NTU)0 to 10, 0 to 100, 0 to 200, 0 to 500, 0 to 1000
Minimum readout value (NTU)0.01
Calibration point5
Portable Turbidity Meter LTM-A15

Portable Turbidity Meter LTM-A15

Measuring range (NTU)0 to 100, 100 to 1000
Minimum readout value (NTU)0.01, 0.1
Calibration point7
Portable Turbidity Meter LTM-A14

Portable Turbidity Meter LTM-A14

Measuring range (NTU)0 to 100, 100 to 800
Minimum readout value (NTU)0.01, 0.1
Calibration point6
Bench top Turbidity Meter LTM-B11

Bench top Turbidity Meter LTM-B11

Measuring range (NTU)0 to 50, 0 to 100, 0 to 200
Minimum readout value (NTU)0.01, 0.1
Calibration point3
Portable Turbidity Meter LTM-A11

Portable Turbidity Meter LTM-A11

Measuring range (NTU)0 to 100, 0 to 200
Minimum readout value (NTU)0.01, 0.1
Calibration point4
Portable Turbidity Meter LTM-A13

Portable Turbidity Meter LTM-A13

Measuring range (NTU)0 to 100, 100 to 500
Minimum readout value (NTU)0.01, 0.1
Calibration point6
Bench top Turbidity Meter LTM-B12

Bench top Turbidity Meter LTM-B12

Measuring range (NTU)0 to 50, 0 to 100, 0 to 200, 0 to 400
Minimum readout value (NTU)0.01, 0.1
Calibration point4
Bench top Turbidity Meter LTM-B10

Bench top Turbidity Meter LTM-B10

Measuring range (NTU)0 to 100
Minimum readout value (NTU)0.01, 0.1
Calibration point1
Bench top Turbidity Meter (with built-in printer) LTM-C10

Bench top Turbidity Meter (with built-in printer) LTM-C10

Measuring range (NTU)0 to 100, 0 to 200
Minimum readout value (NTU)0.01
Calibration point2

Online pH Controller

Online pH Controller LOPC-A10

Online pH Controller LOPC-A10

pH measurement range01.00 to 15.00 pH
pH calibration points1 to 3 points
pH Buffer SolutionsUSA (pH 4.01, 7.00, 10.01) or NIST (pH 4.01, 6.86, 9.18)

Online Dissolved Oxygen Controller

Online Dissolved Oxygen Controller LODC-A10

Online Dissolved Oxygen Controller LODC-A10

Dissolved oxygen range0.0 to 20.0 mg / L (or ppm)
Dissolved oxygen Calibration points1 - 2 points
Dissolved oxygen accuracy± 0.5 mg / L

Online Conductivity Controller

Online Conductivity Controller LOCC-A10

Online Conductivity Controller LOCC-A10

Conductivity measurement range0.01 to 20.00, 200.0, 2000 µS/ cm, 20.00, 200.0 mS/ cm
Conductivity calibration points1 to 3 points
Conductivity accuracy±1 % F.S.

TSS Meter

TSS Meter LTSS-A10

TSS Meter LTSS-A10

Measurement range20000 mg/L <br/> 0.01 mg/L ~ 45000 mg/L <br/> 0.01 mg/L ~ 120000 mg/L
Measurement accuracy± 2 % F.S. of its range ( Depending on the homogeneity of sludge)
Pressure range≤ 0.4 Mpa

Benchtop Multi-parameter Water Quality Meter

Benchtop Multi-parameter Water Quality Meter LMPWM-A30

Benchtop Multi-parameter Water Quality Meter LMPWM-A30

pH Mode
pH range -2.00 to 20.00 pH
pH Accuracy ± 0.002 pH
pH Resolution 0.1/0.01/0.001 pH, selectable
pH Calibration Points 1 to 5 Points
pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
mV Mode
mV Range -1999.9 to 1999.9 mV
mV Accuracy ± 0.2 mV
mV Resolution 0.1/1 mV, selectable
mV Calibration Points 1 Point
Benchtop Multi-parameter Water Quality Meter LMPWM-A31

Benchtop Multi-parameter Water Quality Meter LMPWM-A31

pH Mode
pH range -2.00 to 20.00 pH
pH Accuracy ± 0.002 pH
pH Resolution 0.1/0.01/0.001 pH, selectable
pH Calibration Points 1 to 5 Points
pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
mV Mode
mV Range -1999.9 to 1999.9 mV
mV Accuracy ± 0.2 mV
mV Resolution 0.1/1 mV, selectable
mV Calibration Points 1 Point
Benchtop Multi-parameter Water Quality Meter LMPWM-A40

Benchtop Multi-parameter Water Quality Meter LMPWM-A40

pH Mode
pH range -2.00 to 20.00 pH
pH Accuracy ± 0.002 pH
pH Resolution 0.1/0.01/0.001 pH, selectable
pH Calibration Points 1 to 5 Points
pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
ORP Mode
mV Range ± 2000 mV
mV Accuracy ± 0.2 mV
mV Resolution 0.1 mV, selectable
mV Calibration Points 1 Point, only for relative mode
Benchtop Multi-parameter Water Quality Meter LMPWM-A32

Benchtop Multi-parameter Water Quality Meter LMPWM-A32

pH Mode
pH range -2.00 to 20.00 pH
pH Accuracy ± 0.002 pH
pH Resolution 0.1/0.01/0.001 pH, selectable
pH Calibration Points 1 to 5 Points
pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
mV Mode
mV Range -1999.9 to 1999.9 mV
mV Accuracy ± 0.2 mV
mV Resolution 0.1/1 mV, selectable
mV Calibration Points 1 Point

BOD Meter

BOD Meter LBOD-A10

BOD Meter LBOD-A10

Determination range
0 to 35 mg/L, 0 to 70 mg/ L, 0 to 350 mg/L, 0 to 700 mg/L (without dilution)
0 to 4000 mg/ L (need to dilute) (subsection measurement)
Measurement errorLess or less than 10 %
Sample container/ bottle quantity1 to 6

Portable Conductivity Meter

Portable Conductivity Meter LPCM-A11

Portable Conductivity Meter LPCM-A11

Conductivity
Conductivity measurement range: 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200 mS/cm
Conductivity accuracy: ±0.5% F.S
Conductivity calibration points: 1 to 5 points
Conductivity calibration solutions: 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm or Custom
Conductivity resolution: 0.001, 0.01, 0.1, 1
TDS
TDS range: 0 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
TDS accuracy: ±1 % F.S
TDS factor: 0.01 to 1.00 (Default 0.5)
TDS resolution: 0.01, 0.1, 1
Temperature
Temperature measuring range: 0 105 °C, 32 to 221 °F
Temperature measuring accuracy: ± °0.5 C, ± 0.9 °F
Temperature calibration points: 1 point
Temperature resolution: 0.1 °C, 0.1 °F
Portable Conductivity Meter LPCM-A12

Portable Conductivity Meter LPCM-A12

Conductivity
Conductivity measurement range: 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200 mS/cm
Conductivity accuracy: ±0.5% F.S
Conductivity calibration points: 1 to 5 points
Conductivity calibration solutions: 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm or Custom
Conductivity resolution: 0.001, 0.01, 0.1, 1
Salinity
Salinity range: 0.00 to 80.00 ppt, 0.00 to 42.00 psu
Salinity accuracy: ±1% F.S
Salinity measurement modes: Practical Salinity (psu) or Natural Seawater (ppt)
Salinity resolution: 0.01, 0.1, 1
Temperature
Temperature measuring range: 0 105 °C, 32 to 221 °F
Temperature measuring accuracy: ± °0.5 C, ± 0.9 °F
Temperature calibration points: 1 point
Temperature resolution: 0.1 °C, 0.1 °F
Portable Conductivity Meter LPCM-A13

Portable Conductivity Meter LPCM-A13

Conductivity
Conductivity measurement range: 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200 mS/cm
Conductivity accuracy: ±0.5% F.S
Conductivity calibration points: 1 to 5 points
Conductivity calibration solutions: 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm or Custom
Conductivity resolution: 0.001, 0.01, 0.1, 1
TDS
TDS range: 0 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
TDS accuracy: ±1 % F.S
TDS factor: 0.01 to 1.00 (Default 0.5)
TDS resolution: 0.01, 0.1, 1
Salinity
Salinity range: 0.00 to 80.00 ppt, 0.00 to 42.00 psu
Salinity accuracy: ±1% F.S
Salinity measurement modes: Practical Salinity (psu) or Natural Seawater (ppt)
Salinity resolution: 0.01, 0.1, 1
Portable Conductivity Meter LPCM-A10

Portable Conductivity Meter LPCM-A10

Conductivity
Conductivity measurement range: 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200 mS/cm
Conductivity accuracy: ±0.5% F.S
Conductivity calibration points: 1 to 3 points
Conductivity calibration solutions: 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm or Custom
Conductivity resolution: 0.001, 0.01, 0.1, 1
Temperature
Temperature measuring range: 0 105 °C, 32 to 221 °F
Temperature measuring accuracy: ± °0.5 C, ± 0.9 °F
Temperature calibration points: 1 point
Temperature resolution: 0.1 °C, 0.1 °F
Temperature compensation0 to 100 °C, 32 to 212 ° F, Manual or Automatic

Water Activity Meter

Water Activity Meter LWTM-A10

Water Activity Meter LWTM-A10

Range0 to 1.0 aw
Resolution0 to 1.0 aw
Accuracy± 0.02 aw
Water Activity Meter with sensor LWAM-A10

Water Activity Meter with sensor LWAM-A10

Measuring rangeWater activity 0 to 0.980 aw (no condensing )
Temperature range0 to 50°C
Measurement accuracy± 0.015 aw

Portable Chlorine Meter

Portable Chlorine Meter LPCL-A10

Portable Chlorine Meter LPCL-A10

Residual chlorine/total chlorine
Range: 0.00 to 5.00 mg/L
Accuracy: 0.001 mg/L 0.001 ABS
Repeatability Lower detection limit: ≤0.02 mg/L
Chlorine dioxide
Range: 0.00 to 10.00 mg/L
Resolution: 0.001 mg/L 0.001 ABS
Accuracy: ≤2 mg/L: ±0.10 mg/L ˃2 mg/L: ±5 %
Lower detection limit: ≤0.04 mg/L
Light sourceLED