Online Conductivity Controller LOCC-A10

Online Conductivity Controller LOCC-A10 features 3 points conductivity calibration system, TDS measurement mode, and 4 to 20 mA output signal transmitter. The large and clear back light LCD displays Calibration and setup values, Temperature values, ATC (automatic temperature compensation) values, high and low limit of measurement. This controller measures Conductivity and TDS values in µS/ cm, and mS/ cm units.

Conductivity measurement range 0.01 to 20.00, 200.0, 2000 µS/ cm, 20.00, 200.0 mS/ cm
Conductivity calibration points 1 to 3 points
Conductivity accuracy ±1 % F.S.
Conductivity resolution 0.001/ 0.01/ 0.1/ 1
Conductivity calibration solutions 84 µS/ cm, 1413 µS/ cm, 12.88 mS/ cm,111.8 mS/ cm
TDS measurement range 0 to 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
TDS accuracy ±1 % F.S.
TDS factor 0.1 to 1.0
Temperature measurement range 0 to 100°C, 32 to 212°F
Temperature accuracy 1°C
Temperature resolution 0.1°C
Temperature calibration points 1 point
Temperature calibration range Measured value ± 10°C
Signal output 4 to 200 mA
Load Max. 500 Ω
Low and high alarm limits 0.02 uS/ cm to 20.0 mS/ cm, selectable
Interface RS485
Connection terminals Detachable screw terminals
Temperature compensation Automatic, 0 to 100°C , 32 to 212°F
Cell constant K= 0.1, 1, 10
Environmental conditions Ambient temperature to ≤ 60°C, RH ≤ 80%
Reference temperature 25°C
Display LCD
Power DC 24 V
Dimension (L*W*H) 96 × 96 × 75 mm
Weight 350 g
  • Equipped with industrial conductivity electrode
  • 1 to 3 conductivity calibration points, 1 temperature calibration point
  • Selectable temperature units °C or °F for diff. applications
  • System menu allows setting of up to 9 parameters
  • Reset features to resumes all setting back to default
  • ATC mode for accurate readings over entire range
  • Low and high alarm limits set automatically activate an external device
  • Manual temperature calibration improves accuracy
Accessories no. Name Qty.
1 Industrial Conductivity electrode for measuring pure water 1
2 Industrial Conductivity electrode for measuring general liquids 1
3 Industrial Conductivity electrode for measuring high conductivity liquids 1
4 Conductivity calibration solutions (bottle of 84 µS/ cm, 1413 µS/ cm, 12.88 mS/ cm,111.8 mS/ cm) 1 each

Used to measure conductivity and TDS values required for quality checking and research purpose across water treatment plants, filtration units, RO treatment plants, Agriculture industry, Aquaculture industry and others.

Benchtop Ion Meter

Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.

Ion measurement range Cl range: 5 × 10<sup>6</sup> - 1 M, 1.8 - 35500 ppm
Ion measurement accuracy ± 0.5 % F.S.
Resolution 0.001, 0.01, 0.1, 1
Calibration points 2 to 5 points
Ion measurement range Na range: 1 × 10<sup>5</sup> - 1 M, 0.1 - 23000 ppm
Ion measurement accuracy ± 0.5 % F.S.
Resolution 0.001, 0.01, 0.1, 1
Calibration points 2 to 5 points
Ion measurement range F range: 1 × 10<sup>6</sup>M – saturation, 0.02 ppm to saturation
Ion measurement accuracy ± 0.5 % F.S.
Resolution 0.001, 0.01, 0.1, 1
Calibration points 2 to 5 points

Dissolved Oxygen Tester

Labtron pocket testers are small and compact in size, gives fast and accurate results. The Pocket testers consist of pH, ORP, TDS, dissolved oxygen, salinity and conductivity tester for various laboratory as well as outdoor applications. Measurements can be taken by inserting the probe in to the test solution and after gentle stirring, stable readings can be noted down.

Dissolved oxygen range 0.0 to 20.0 mg/L (or ppm)
Dissolved oxygen accuracy ± 0.5 mg/L
% of saturation of oxygen 0.0 to 200.0%
Sensor type Polarographic dissolved oxygen probe

Salinity Tester

Labtron pocket testers are small and compact in size, gives fast and accurate results. The Pocket testers consist of pH, ORP, TDS, dissolved oxygen, salinity and conductivity tester for various laboratory as well as outdoor applications. Measurements can be taken by inserting the probe in to the test solution and after gentle stirring, stable readings can be noted down.

Salinity measurement range 0.00 to 80.00 ppt
Salinity accuracy ± 1 % F.S
Conductivity range 0 - 200.0 µS/cm, 100 - 2000 µS/cm, 20.00 mS/cm
Calibration solutions 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm
Salinity measurement range 0.00 to 10.00 ppt
Salinity accuracy ± 1 % F.S
Conductivity range 0 - 20.00, 200.0, 2000 µS/cm, 20.00 mS/cm
Calibration solutions 84 µS/cm, 1413 µS/cm, 12.88 mS/cm

pH Meters (Portable)

Labtron portable meters are used in quality analysis of water and other solutions. With portability, waterproof features, finds applications in indoor as well as outdoor measurement of pH, ORP, dissolved oxygen, water hardness, salinity, TDS etc.

pH range (-2.00) to 16.00 pH
pH accuracy ± 0.01 pH
pH resolution 0.01 pH
Calibration points 2 points
pH range (-2.00) to 20.00 pH
pH accuracy ± 0.002 pH
pH resolution 0.1, 0.01, 0.001 pH
Calibration points 1 - 5 points
Measuring range (NTU) 0 to 100, 0 to 200
Minimum readout value (NTU) 0.01, 0.1
Calibration point 4
Sample volume 5 to 10 ml
Measuring range (NTU) 0 to 200
Minimum readout value (NTU) 0.1
Calibration point Nil
Sample volume 5 to 10 ml
Light source LED/860 nm (Wavelength)
Readings display LCD
Unit switches NTU, FTU, EBC, ASBC
Measurement range 0.00-20.0; 20.01-200; 200.01-1000 NTU