Pocket Salinity tester LPST-A11

Pocket Salinity tester LPST-A11 comes with 1 - 3 point calibration system. Equipped with platinum sensor, it provides fast and reliable results. The System menu allows setting of up to 5 parameters. It measures salinity of water and other low and high salt concentrated solutions. Equipped with replaceable electrode helps to reduce the maintenance cost.

Salinity measurement range 0.00 to 80.00 ppt
Salinity accuracy ± 1 % F.S
Conductivity range 0 - 200.0 µS/cm, 100 - 2000 µS/cm, 20.00 mS/cm
Calibration solutions 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm
Conductivity accuracy ± 1 % F.S
Calibration points 1 - 3 points
Temperature compensation range 0 °C to 60 °C, 32 °F to 140 °F
Temperature accuracy ± 1 °C
Temperature coefficient 2 % per °C
Normalization temperature 25 °C
Reset function Yes
Cell constant K=1
Data hold function Manual or automatic endpoint detection
Power off Manual or automatic (8 minute after last key is pressed)
Sensor type Platinum sensor
Battery life Approximately 200 hours
Power 2 × 1.5 V AAA batteries
Dimension (L × Dia.) 185 × 40 mm
Weight 100 g
  • Low to high measurement range of salinity
  • Equipped with platinum sensor
  • 1 - 3 points push button calibration with automatic recognition of calibration solutions
  • Automatic temperature compensation for accurate measured value over the entire range
  • Auto hold function freezes stable, final readings for better viewing and recording
  • Automatic power off helps to reduce battery power consumption
  • Waterproof features guarantees protection against harsh conditions
  • Battery indicator to show the remaining battery capacity
  • Reset function automatically resumes all the settings to default mode
  • System menu allows setting of up to 5 parameters
1413µS/cm, 12.88mS/cm , 111.8mS/cm calibration standard solutions
Platinum conductivity electrode

Pocket salinity tester measures salinity i.e. dissolved salt from pure water, brine water, aquarium water, food & beverage processing units & other industrial applications.

Related Products of Salinity Tester

Labtron pocket testers are small and compact in size, gives fast and accurate results. The Pocket testers consist of pH, ORP, TDS, dissolved oxygen, salinity and conductivity tester for various laboratory as well as outdoor applications. Measurements can be taken by inserting the probe in to the test solution and after gentle stirring, stable readings can be noted down.

Salinity measurement range 0.00 to 10.00 ppt
Salinity accuracy ± 1 % F.S
Conductivity range 0 - 20.00, 200.0, 2000 µS/cm, 20.00 mS/cm
Calibration solutions 84 µS/cm, 1413 µS/cm, 12.88 mS/cm

Portable Water Hardness Meters

Labtron portable meters are used in quality analysis of water and other solutions. With portability, waterproof features, finds applications in indoor as well as outdoor measurement of pH, ORP, dissolved oxygen, water hardness, salinity, TDS etc.

Water hardness range 0.05 mmol / L - 0.2 mol / L<br/> (CaCO<sub>3</sub>) 0 - 19999 mg / L, (CaO) 0 - 11220 mg / L, (boiler) 0 - 400 mmol / L, (Ca) 0 - 8020 mg / L<br/> (German degree) 0 - 1122 °dH, (English degree) 0 - 1404 °e, (French degree) 0 - 2000 °f
Accuracy ± 1 % F.S.
Calibration points 2 to 5 points
Calibration solution 0.001, 0.01, 0.1 mmol / L, 0.001, 0.01, 0.1 mol / L

Benchtop Ion Meter

Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.

Ion measurement range F range: 1 × 10<sup>6</sup>M – saturation, 0.02 ppm to saturation
Ion measurement accuracy ± 0.5 % F.S.
Resolution 0.001, 0.01, 0.1, 1
Calibration points 2 to 5 points
Ion measurement range Cl range: 5 × 10<sup>6</sup> - 1 M, 1.8 - 35500 ppm
Ion measurement accuracy ± 0.5 % F.S.
Resolution 0.001, 0.01, 0.1, 1
Calibration points 2 to 5 points
Ion measurement range Na range: 1 × 10<sup>5</sup> - 1 M, 0.1 - 23000 ppm
Ion measurement accuracy ± 0.5 % F.S.
Resolution 0.001, 0.01, 0.1, 1
Calibration points 2 to 5 points

pH Meters (Portable)

Labtron portable meters are used in quality analysis of water and other solutions. With portability, waterproof features, finds applications in indoor as well as outdoor measurement of pH, ORP, dissolved oxygen, water hardness, salinity, TDS etc.

pH range (-2.00) to 20.00 pH
pH accuracy ± 0.002 pH
pH resolution 0.1, 0.01, 0.001 pH
Calibration points 1 - 5 points
pH range (-2.00) to 16.00 pH
pH accuracy ± 0.01 pH
pH resolution 0.01 pH
Calibration points 2 points
COD
COD determination range 5 to 5000 mg/ L
COD measurement accuracy COD < 50 mg/ L, ≤ ± 10 % and COD > 50 mg/ L, ≤ ± 5 %
COD minimum detection limit 0.1 mg/ L
COD measurement time 20 minutes
COD repeatability ≤ ± 5 %
Processing sample number per batch 12 pcs
Consumables for colorimetric COD measurement Colorimetric tubes and Cuvette
COD reactor digestion temperature 165 ± 0.5°C
COD reactor digestion time 10 min
COD reactor time setting switches 3 pcs
Anti-chlorine interference (Cl) < 1000 mg/ L No effect
Ammonia nitrogen
Ammonia nitrogen determination range 0.02 to 30 mg/ L
Ammonia nitrogen measurement accuracy ≤ ± 10 %
Ammonia nitrogen minimum detection limit 0.01 mg/ L
Ammonia nitrogen measurement time 10 to 15 minutes
Ammonia nitrogen repeatability ≤ ± 5 %
Processing sample number per batch 20 pcs
Consumables for colorimetric ammonia nitrogen measurement Colorimetric tubes and Cuvette
Ammonia nitrogen reactor time setting switches 3 pcs
Total phosphorous
Total phosphorous determination range 0.002 to 10 mg/ L
Total phosphorous total phosphorous measurement accuracy ≤ ± 5 %
Total phosphorous minimum detection limit 0.001 mg/ L
Total phosphorous measurement time 35 to 50 minutes
Total phosphorous repeatability ≤ ± 5 %
Processing sample number per batch 12 pcs
Consumables for colorimetric total phosphorous measurement Colorimetric tubes and Cuvette
Total phosphorous reactor digestion temperature 120 ± 0.5°C
Total phosphorous reactor digestion time 30 min
Total phosphorous reactor time setting switches 3 pcs
Turbidity
Turbidity determination range 0.5 to 400 NTU
Turbidity measurement accuracy ≤ ± 5 %
Turbidity minimum detection limit 0.1 NTU
Turbidity measurement time 1 minute
Turbidity repeatability ≤ ± 5 %
Consumables for colorimetric turbidity measurement Colorimetric Cuvette
COD
COD determination range 2 to 10000 mg/ L
COD measurement accuracy COD < 50 mg/ L, ≤ ± 10 % and COD > 50 mg/ L, ≤ ± 5 %
COD minimum detection limit 0.1 mg/ L
COD measurement time 20 minutes
COD repeatability ≤ ± 5 %
Processing sample number per batch
Consumables for colorimetric COD measurement Colorimetric tubes and Cuvette
COD reactor digestion temperature 165 ± 0.5°C
COD reactor digestion time 10 min
COD reactor time setting switches 3 pcs
Anti-chlorine interference (Cl) < 1000 mg/ L No effect
Ammonia nitrogen
Ammonia nitrogen determination range 0.02 to 60 mg/ L
Ammonia nitrogen measurement accuracy ≤ ± 10 %
Ammonia nitrogen minimum detection limit 0.01 mg/ L
Ammonia nitrogen measurement time 10 to 15 minutes
Ammonia nitrogen repeatability ≤ ± 5 %
Processing sample number per batch
Consumables for colorimetric ammonia nitrogen measurement Colorimetric tubes and Cuvette
Ammonia nitrogen reactor time setting switches 3 pcs
Total phosphorous
Total phosphorous determination range 0.002 to 20 mg/ L
Total phosphorous total phosphorous measurement accuracy ≤ ± 5 %
Total phosphorous minimum detection limit 0.001 mg/ L
Total phosphorous measurement time 35 to 50 minutes
Total phosphorous repeatability ≤ ± 5 %
Processing sample number per batch
Consumables for colorimetric total phosphorous measurement Colorimetric tubes and Cuvette
Total phosphorous reactor digestion temperature 120 ± 0.5°C
Total phosphorous reactor digestion time 30 min
Total phosphorous reactor time setting switches 3 pcs
Turbidity
Turbidity determination range 0.5 to 400 NTU
Turbidity measurement accuracy ≤ ± 5 %
Turbidity minimum detection limit 0.1 NTU
Turbidity measurement time 1 minute
Turbidity repeatability ≤ ± 5 %
Consumables for colorimetric turbidity measurement Colorimetric Cuvette
Measurement range 20000 mg/L <br/> 0.01 mg/L ~ 45000 mg/L <br/> 0.01 mg/L ~ 120000 mg/L
Measurement accuracy ± 2 % F.S. of its range ( Depending on the homogeneity of sludge)
Pressure range ≤ 0.4 Mpa
Flow rate ≤ 2.5 m/s, 8.2 ft./s