Benchtop Conductivity Meter LBCM-A14

Benchtop Conductivity Meter LBCM-A14 with 1 to 3 points push button conductivity calibration, automatically recognizes calibration standards. With setup menu, it does setting of 6 parameters calibration points, temperature units and others. Large LCD display shows settings and measurement values. Temperature calibration by manual method improves the accuracy of conductivity meter.

Conductivity Mode
  • Conductivity measurement range: 0.01-20.0, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
  • Conductivity accuracy: ±1% F.S.
  • Conductivity resolution: 0.001, 0.01, 0.1, 1
  • Conductivity calibration points: 1 to 3 points
  • Conductivity calibration solutions: 10 µS/cm , 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm
TDS Mode
  • TDS measurement Range: 0 to 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
  • TDS Accuracy: ± 1 % F.S.
  • TDS Factor: 0.01 to 1.00 (default 0.5)
  • TDS Calibration points: 3 points
Temperature Mode
  • Temperature range: 0 to 105°C, 32 to 221°F
  • Temperature accuracy: ±1°C
  • Temperature resolution: 0.1°C, 0.1°F
  • Temperature calibration points: 1 point
  • Temperature Compensation Range and Mode: Manual or Automatic, 0 to 100°C , 32 to 212°F
  • Temperature coefficient: 0.0 to 10.0 % per °C
Cell constant K=0.1, 1, 10
Power requirements DC 9 V, using AC adapter, AC 220 V, 50 Hz
Dimension (L×W×H) 210 × 205 × 75 mm
Weight 1.5 kg
  • Measures conductivity, TDS, Temperature of the liquids
  • Automatic temperature compensation for accurate values over the entire range
  • Selectable temperature units (°C or °F) for different application requirements
  • Reset function automatically resumes all the setting back to factory default options
  • Manual and automatic mode of temperature calibration
  • Auto hold feature to lock, record and read stable measurement endpoint
  • Stability indicator displays the status of current measurement
  • Help message as an operational guide, with quick access to test parameters
  • Conductivity/TDS Mode
  • 1 to 3 Conductivity push button calibration points
  • 3 TDS calibration points
  • 1 Temperature calibration point
Accessories no. Name
1 Conductivity Electrode (K=0.1) suitable for measuring the low conductivity liquids ( <10µS/ cm) e.g. Pure water
2 Conductivity Electrode (K=10) suitable for measuring the high conductivity liquids (>20mS/ cm) e.g. Sea water

Used to test Conductivity, TDS values of different type of liquid such as pure, hard/low conductivity water, distilled water. Also used in water purification labs, water testing and management plants, and water processing and packaging units for quality and research purpose

Related Products of Benchtop Conductivity Meter

Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.

Conductivity measurement range 0 - 20.0, 200.0, 2000 µS / cm, 200.0 mS / cm
Conductivity accuracy ± 1 % F.S.
Conductivity resolution 0.01, 0.1, 1
Conductivity calibration solutions 10 µS / cm , 84 µS / cm, 1413 µS / cm, 12.88 mS / cm, 111.8 mS / cm
Conductivity measurement range 0.000 μS/cm to 3000mS/cm
Resistivity measurement range 5.00 Ω·cm to 100.0 MΩ·cm
TDS measurement range 0.000 mg/L to 1000g/L
Salinity measurement range 0.00 to 8.00 %
pH measurement range 01.00 to 15.00 pH
pH calibration points 1 to 3 points
pH Buffer Solutions USA (pH 4.01, 7.00, 10.01) or NIST (pH 4.01, 6.86, 9.18)
pH accuracy ± 0.01 pH

ORP Tester

Labtron pocket testers are small and compact in size, gives fast and accurate results. The Pocket testers consist of pH, ORP, TDS, dissolved oxygen, salinity and conductivity tester for various laboratory as well as outdoor applications. Measurements can be taken by inserting the probe in to the test solution and after gentle stirring, stable readings can be noted down.

Connector BNC
ORP measurement range (-999) to 999 mV
ORP accuracy ± 2 mV
ORP resolution 1 mV
Sensor type Platinum sensor
Connector 6 pin
ORP measurement range (-999) to 999 mV
ORP accuracy ± 2 mV
COD
COD determination range 2 to 10000 mg/ L
COD measurement accuracy COD < 50 mg/ L, ≤ ± 10 % and COD > 50 mg/ L, ≤ ± 5 %
COD minimum detection limit 0.1 mg/ L
COD measurement time 20 minutes
COD repeatability ≤ ± 5 %
Processing sample number per batch
Consumables for colorimetric COD measurement Colorimetric tubes and Cuvette
COD reactor digestion temperature 165 ± 0.5°C
COD reactor digestion time 10 min
COD reactor time setting switches 3 pcs
Anti-chlorine interference (Cl) < 1000 mg/ L No effect
Ammonia nitrogen
Ammonia nitrogen determination range 0.02 to 60 mg/ L
Ammonia nitrogen measurement accuracy ≤ ± 10 %
Ammonia nitrogen minimum detection limit 0.01 mg/ L
Ammonia nitrogen measurement time 10 to 15 minutes
Ammonia nitrogen repeatability ≤ ± 5 %
Processing sample number per batch
Consumables for colorimetric ammonia nitrogen measurement Colorimetric tubes and Cuvette
Ammonia nitrogen reactor time setting switches 3 pcs
Total phosphorous
Total phosphorous determination range 0.002 to 20 mg/ L
Total phosphorous total phosphorous measurement accuracy ≤ ± 5 %
Total phosphorous minimum detection limit 0.001 mg/ L
Total phosphorous measurement time 35 to 50 minutes
Total phosphorous repeatability ≤ ± 5 %
Processing sample number per batch
Consumables for colorimetric total phosphorous measurement Colorimetric tubes and Cuvette
Total phosphorous reactor digestion temperature 120 ± 0.5°C
Total phosphorous reactor digestion time 30 min
Total phosphorous reactor time setting switches 3 pcs
Turbidity
Turbidity determination range 0.5 to 400 NTU
Turbidity measurement accuracy ≤ ± 5 %
Turbidity minimum detection limit 0.1 NTU
Turbidity measurement time 1 minute
Turbidity repeatability ≤ ± 5 %
Consumables for colorimetric turbidity measurement Colorimetric Cuvette
COD
COD determination range 5 to 5000 mg/ L
COD measurement accuracy COD < 50 mg/ L, ≤ ± 10 % and COD > 50 mg/ L, ≤ ± 5 %
COD minimum detection limit 0.1 mg/ L
COD measurement time 20 minutes
COD repeatability ≤ ± 5 %
Processing sample number per batch 12 pcs
Consumables for colorimetric COD measurement Colorimetric tubes and Cuvette
COD reactor digestion temperature 165 ± 0.5°C
COD reactor digestion time 10 min
COD reactor time setting switches 3 pcs
Anti-chlorine interference (Cl) < 1000 mg/ L No effect
Ammonia nitrogen
Ammonia nitrogen determination range 0.02 to 30 mg/ L
Ammonia nitrogen measurement accuracy ≤ ± 10 %
Ammonia nitrogen minimum detection limit 0.01 mg/ L
Ammonia nitrogen measurement time 10 to 15 minutes
Ammonia nitrogen repeatability ≤ ± 5 %
Processing sample number per batch 20 pcs
Consumables for colorimetric ammonia nitrogen measurement Colorimetric tubes and Cuvette
Ammonia nitrogen reactor time setting switches 3 pcs
Total phosphorous
Total phosphorous determination range 0.002 to 10 mg/ L
Total phosphorous total phosphorous measurement accuracy ≤ ± 5 %
Total phosphorous minimum detection limit 0.001 mg/ L
Total phosphorous measurement time 35 to 50 minutes
Total phosphorous repeatability ≤ ± 5 %
Processing sample number per batch 12 pcs
Consumables for colorimetric total phosphorous measurement Colorimetric tubes and Cuvette
Total phosphorous reactor digestion temperature 120 ± 0.5°C
Total phosphorous reactor digestion time 30 min
Total phosphorous reactor time setting switches 3 pcs
Turbidity
Turbidity determination range 0.5 to 400 NTU
Turbidity measurement accuracy ≤ ± 5 %
Turbidity minimum detection limit 0.1 NTU
Turbidity measurement time 1 minute
Turbidity repeatability ≤ ± 5 %
Consumables for colorimetric turbidity measurement Colorimetric Cuvette

Benchtop Dissolved Oxygen Meter

Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.

Dissolved oxygen range 0.0 to 20.0 mg / L (or ppm)
Dissolved oxygen accuracy ± 0.5 mg / L
Resolution 0.1 mg / L
Calibration points 1 - 2 points
Dissolved oxygen range 0.0 to 20.0 mg / L (or ppm)
Dissolved oxygen accuracy ± 0.2 mg / L
Resolution 0.01 mg / L
Calibration points 1 - 2 points
Conductivity
Conductivity measurement range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200 mS/cm
Conductivity accuracy ±0.5% F.S
Conductivity calibration points 1 to 3 points
Conductivity calibration solutions 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm or Custom
Conductivity resolution 0.001, 0.01, 0.1, 1
Temperature
Temperature measuring range 0 105 °C, 32 to 221 °F
Temperature measuring accuracy ± °0.5 C, ± 0.9 °F
Temperature calibration points 1 point
Temperature resolution 0.1 °C, 0.1 °F
Temperature compensation 0 to 100 °C, 32 to 212 ° F, Manual or Automatic
Temperature coefficient Linear
Conductivity
Conductivity measurement range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200 mS/cm
Conductivity accuracy ±0.5% F.S
Conductivity calibration points 1 to 5 points
Conductivity calibration solutions 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm or Custom
Conductivity resolution 0.001, 0.01, 0.1, 1
TDS
TDS range 0 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
TDS accuracy ±1 % F.S
TDS factor 0.01 to 1.00 (Default 0.5)
TDS resolution 0.01, 0.1, 1
Temperature
Temperature measuring range 0 105 °C, 32 to 221 °F
Temperature measuring accuracy ± °0.5 C, ± 0.9 °F
Temperature calibration points 1 point
Temperature resolution 0.1 °C, 0.1 °F
Temperature compensation 0 to 100 °C, 32 to 212 ° F, Manual or Automatic
Conductivity
Conductivity measurement range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200 mS/cm
Conductivity accuracy ±0.5% F.S
Conductivity calibration points 1 to 5 points
Conductivity calibration solutions 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm or Custom
Conductivity resolution 0.001, 0.01, 0.1, 1
Salinity
Salinity range 0.00 to 80.00 ppt, 0.00 to 42.00 psu
Salinity accuracy ±1% F.S
Salinity measurement modes Practical Salinity (psu) or Natural Seawater (ppt)
Salinity resolution 0.01, 0.1, 1
Temperature
Temperature measuring range 0 105 °C, 32 to 221 °F
Temperature measuring accuracy ± °0.5 C, ± 0.9 °F
Temperature calibration points 1 point
Temperature resolution 0.1 °C, 0.1 °F
Temperature compensation 0 to 100 °C, 32 to 212 ° F, Manual or Automatic