Benchtop Conductivity Meter LBCM-A14

Benchtop Conductivity Meter LBCM-A14 with 1 to 3 points push button conductivity calibration, automatically recognizes calibration standards. With setup menu, it does setting of 6 parameters calibration points, temperature units and others. Large LCD display shows settings and measurement values. Temperature calibration by manual method improves the accuracy of conductivity meter.

Conductivity Mode
  • Conductivity measurement range: 0.01-20.0, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
  • Conductivity accuracy: ±1% F.S.
  • Conductivity resolution: 0.001, 0.01, 0.1, 1
  • Conductivity calibration points: 1 to 3 points
  • Conductivity calibration solutions: 10 µS/cm , 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm
TDS Mode
  • TDS measurement Range: 0 to 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
  • TDS Accuracy: ± 1 % F.S.
  • TDS Factor: 0.01 to 1.00 (default 0.5)
  • TDS Calibration points: 3 points
Temperature Mode
  • Temperature range: 0 to 105°C, 32 to 221°F
  • Temperature accuracy: ±1°C
  • Temperature resolution: 0.1°C, 0.1°F
  • Temperature calibration points: 1 point
  • Temperature Compensation Range and Mode: Manual or Automatic, 0 to 100°C , 32 to 212°F
  • Temperature coefficient: 0.0 to 10.0 % per °C
Cell constant K=0.1, 1, 10
Power requirements DC 9 V, using AC adapter, AC 220 V, 50 Hz
Dimension (L×W×H) 210 × 205 × 75 mm
Weight 1.5 kg
  • Measures conductivity, TDS, Temperature of the liquids
  • Automatic temperature compensation for accurate values over the entire range
  • Selectable temperature units (°C or °F) for different application requirements
  • Reset function automatically resumes all the setting back to factory default options
  • Manual and automatic mode of temperature calibration
  • Auto hold feature to lock, record and read stable measurement endpoint
  • Stability indicator displays the status of current measurement
  • Help message as an operational guide, with quick access to test parameters
  • Conductivity/TDS Mode
  • 1 to 3 Conductivity push button calibration points
  • 3 TDS calibration points
  • 1 Temperature calibration point
Accessories no. Name
1 Conductivity Electrode (K=0.1) suitable for measuring the low conductivity liquids ( <10µS/ cm) e.g. Pure water
2 Conductivity Electrode (K=10) suitable for measuring the high conductivity liquids (>20mS/ cm) e.g. Sea water

Used to test Conductivity, TDS values of different type of liquid such as pure, hard/low conductivity water, distilled water. Also used in water purification labs, water testing and management plants, and water processing and packaging units for quality and research purpose

Related Products of Benchtop Conductivity Meter

Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.

Conductivity measurement range 0.000 μS/cm to 3000mS/cm
Resistivity measurement range 5.00 Ω·cm to 100.0 MΩ·cm
TDS measurement range 0.000 mg/L to 1000g/L
Salinity measurement range 0.00 to 8.00 %
Conductivity measurement range 0 - 20.0, 200.0, 2000 µS / cm, 200.0 mS / cm
Conductivity accuracy ± 1 % F.S.
Conductivity resolution 0.01, 0.1, 1
Conductivity calibration solutions 10 µS / cm , 84 µS / cm, 1413 µS / cm, 12.88 mS / cm, 111.8 mS / cm
Dissolved oxygen range 0.0 to 20.0 mg / L (or ppm)
Dissolved oxygen Calibration points 1 - 2 points
Dissolved oxygen accuracy ± 0.5 mg / L
Dissolved oxygen resolution 0.1 mg / L
Residual chlorine/total chlorine
Range 0.00 to 5.00 mg/L
Accuracy 0.001 mg/L 0.001 ABS
Repeatability Lower detection limit ≤0.02 mg/L
Chlorine dioxide
Range 0.00 to 10.00 mg/L
Resolution 0.001 mg/L 0.001 ABS
Accuracy ≤2 mg/L: ±0.10 mg/L ˃2 mg/L: ±5 %
Lower detection limit ≤0.04 mg/L
Light source LED
Wavelength of light 515 nm

Conductivity Tester

Labtron pocket testers are small and compact in size, gives fast and accurate results. The Pocket testers consist of pH, ORP, TDS, dissolved oxygen, salinity and conductivity tester for various laboratory as well as outdoor applications. Measurements can be taken by inserting the probe in to the test solution and after gentle stirring, stable readings can be noted down.

Conductivity measurement range 10 to 1999 µS/cm
Type of conductivity Mid-range conductivity
Conductivity accuracy ± 1 % F.S.
Conductivity calibration points 1 point
Conductivity measurement range 0.1 to 19.99 mS/cm
Type of conductivity High range conductivity
Conductivity accuracy ± 1 % F.S.
Conductivity calibration points 1 point
Conductivity measurement range 1.0 to 199.9 µS/cm
Type of conductivity Low range conductivity
Conductivity accuracy ± 1 % F.S.
Conductivity calibration points 1 point
pH Mode
pH range -2.00 to 20.00 pH
pH Accuracy ± 0.002 pH
pH Resolution 0.1/0.01/0.001 pH, selectable
pH Calibration Points 1 to 5 Points
pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
mV Mode
mV Range -1999.9 to 1999.9 mV
mV Accuracy ± 0.2 mV
mV Resolution 0.1/1 mV, selectable
mV Calibration Points 1 Point
Conductivity Mode
Conductivity Range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
Conductivity Accuracy ± 0.5% F.S.
Conductivity Resolution 0.01/0.1/1
Conductivity Calibration Points 1 to 5 Points
Conductivity Solutions 10, 84, 1413 µS/cm, 12.88, 111.8 mS/ cm
Conductivity Temperature coefficient 0.0 to 10.0 %/°C
Compensation mode Linear or Pure Water
Cell Constant K=0.1, 1, 10 or Custom
Reference temperature 20 or 25°C
TDS Mode
TDS Range 0 to 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
TDS Accuracy ± 1 % F.S.
TDS Factor 0.1  to 1.0 (default 0.5)
pH Mode
pH range -2.00 to 20.00 pH
pH Accuracy ± 0.002 pH
pH Resolution 0.1/0.01/0.001 pH, selectable
pH Calibration Points 1 to 5 Points
pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
mV Mode
mV Range -1999.9 to 1999.9 mV
mV Accuracy ± 0.2 mV
mV Resolution 0.1/1 mV, selectable
mV Calibration Points 1 Point
Conductivity Mode
Conductivity Range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
Conductivity Accuracy ± 0.5% F.S.
Conductivity Resolution 0.01/0.1/1
Conductivity Calibration Points 1 to 5 Points
Conductivity Solutions 10, 84, 1413 µS/cm, 12.88, 111.8 mS/ cm
Conductivity Temperature coefficient 0.0 to 10.0 %/°C
Compensation mode Linear or Pure Water
Cell Constant K=0.1, 1, 10 or Custom
Reference temperature 20 or 25°C
Dissolved Oxygen Mode
Dissolved Oxygen Range 0.00 to 20.00 mg/ L
Dissolved Oxygen Accuracy ± 0.2 mg/ L
Dissolved Oxygen Resolution 0.1/ 0.01 mg/ L, selectable
Dissolved Oxygen Calibration Points 1 or 2 Points
% Saturation Range 0.0 to 200.0 % with ± 2 % accuracy
Pressure correction 60 to 112.5 kPa, 450 to 850 mmHg
Salinity correction 0 to 50 g/ L
pH Mode
pH range -2.00 to 20.00 pH
pH Accuracy ± 0.002 pH
pH Resolution 0.1/0.01/0.001 pH, selectable
pH Calibration Points 1 to 5 Points
pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
mV Mode
mV Range -1999.9 to 1999.9 mV
mV Accuracy ± 0.2 mV
mV Resolution 0.1/1 mV, selectable
mV Calibration Points 1 Point
Conductivity Mode
Conductivity Range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
Conductivity Accuracy ± 0.5% F.S.
Conductivity Resolution 0.01/0.1/1
Conductivity Calibration Points 1 to 5 Points
Conductivity Solutions 10, 84, 1413 µS/cm, 12.88, 111.8 mS/ cm
Conductivity Temperature coefficient 0.0 to 10.0 %/°C
Compensation mode Linear or Pure Water
Cell Constant K=0.1, 1, 10 or Custom
Reference temperature 20 or 25°C
TDS Mode
TDS Range 0 to 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
TDS Accuracy ± 1 % F.S.
TDS Factor 0.1  to 1.0 (default 0.5)

TDS Tester

Labtron pocket testers are small and compact in size, gives fast and accurate results. The Pocket testers consist of pH, ORP, TDS, dissolved oxygen, salinity and conductivity tester for various laboratory as well as outdoor applications. Measurements can be taken by inserting the probe in to the test solution and after gentle stirring, stable readings can be noted down.

TDS range 0.5 - 100.0 ppm
TDS accuracy ± 1 % F.S.
TDS factor 0.4 to 1.0 (default 0.5)
Calibration points 1 point
TDS range 5 - 1000 ppm
TDS accuracy ± 1 % F.S.
TDS factor 0.4 to 1.0 (default 0.5)
Calibration points 1 point
TDS range 0.05 - 10.00 ppt
TDS accuracy ± 1 % F.S.
TDS factor 0.4 to 1.0 (default 0.5)
Calibration points 1 point