Portable water hardness meter LPWM-A10

Portable water hardness meter LPWM-A10 comes with LCD display, 2 to 5 points calibration system with low to high concentration range. It can measure salts of calcium, magnesium at mg/L, mmol/ scale. The calibration due reminder alerts user for calibration of the meter on regular basis. The System menu allows setting the 7 parameters like stability condition, auto-power off etc.

Water hardness range 0.05 mmol / L - 0.2 mol / L<br/> (CaCO<sub>3</sub>) 0 - 19999 mg / L, (CaO) 0 - 11220 mg / L, (boiler) 0 - 400 mmol / L, (Ca) 0 - 8020 mg / L<br/> (German degree) 0 - 1122 °dH, (English degree) 0 - 1404 °e, (French degree) 0 - 2000 °f
Accuracy ± 1 % F.S.
Calibration points 2 to 5 points
Calibration solution 0.001, 0.01, 0.1 mmol / L, 0.001, 0.01, 0.1 mol / L
Temperature compensation range 0 - 50 °C
Compensation mode Manual or automatic
Data hold function Manual or automatic endpoint detection
Reset function Yes
Stability condition Low or high
Calibration due 0 to 31 days
Memory Stores up to 500 sets of data
Output USB communication interface
Connector BNC
Battery life Approximately 150 hours
Power 3 × 1.5 V “AA” batteries
Dimension (L × W × H) 170 × 85 × 30 mm
Weight 300 g
  • Equipped with Large LCD display
  • 2 to 5 points calibration with low to high range
  • System menu with 7 parameters setting
  • Calibration due reminder for regular basis calibration of the meter
  • Stability indicator to automatically display current measurements
  • Manual temperature compensation for better measurement accuracy of meter
  • Auto hold function freezes stable, final readings for better viewing and recording
  • Automatic temperature compensation for accurate measured value over the entire range
  • Reset function automatically resumes all the settings to default setting
  • Expanded memory can store up to 500 sets of data
  • Built-in real time clock stamps stored data to meet GLP standards
Water hardness electrode range 0.00005 to 0.2 M

Portable water hardness meter is used in testing the hardness level of tap water, drinking water, filtered water, water used for different experiments and testing across food, textile, painting & other industries.

Measurement range 20000 mg/L <br/> 0.01 mg/L ~ 45000 mg/L <br/> 0.01 mg/L ~ 120000 mg/L
Measurement accuracy ± 2 % F.S. of its range ( Depending on the homogeneity of sludge)
Pressure range ≤ 0.4 Mpa
Flow rate ≤ 2.5 m/s, 8.2 ft./s

Benchtop Conductivity Meter

Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.

Conductivity measurement range 0.000 μS/cm to 3000mS/cm
Resistivity measurement range 5.00 Ω·cm to 100.0 MΩ·cm
TDS measurement range 0.000 mg/L to 1000g/L
Salinity measurement range 0.00 to 8.00 %
Conductivity Mode
Conductivity measurement range 0.01-20.0, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
Conductivity accuracy ±1% F.S.
Conductivity resolution 0.001, 0.01, 0.1, 1
Conductivity calibration points 1 to 3 points
Conductivity calibration solutions 10 µS/cm , 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm
TDS Mode
TDS measurement Range 0 to 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
TDS Accuracy ± 1 % F.S.
TDS Factor 0.01 to 1.00 (default 0.5)
TDS Calibration points 3 points
Temperature Mode
Temperature range 0 to 105°C, 32 to 221°F
Temperature accuracy ±1°C
Temperature resolution 0.1°C, 0.1°F
Temperature calibration points 1 point
Temperature Compensation Range and Mode Manual or Automatic, 0 to 100°C , 32 to 212°F
Temperature coefficient 0.0 to 10.0 % per °C
Cell constant K=0.1, 1, 10
Conductivity measurement range 0 - 20.0, 200.0, 2000 µS / cm, 200.0 mS / cm
Conductivity accuracy ± 1 % F.S.
Conductivity resolution 0.01, 0.1, 1
Conductivity calibration solutions 10 µS / cm , 84 µS / cm, 1413 µS / cm, 12.88 mS / cm, 111.8 mS / cm
pH measurement range 01.00 to 15.00 pH
pH calibration points 1 to 3 points
pH Buffer Solutions USA (pH 4.01, 7.00, 10.01) or NIST (pH 4.01, 6.86, 9.18)
pH accuracy ± 0.01 pH
Conductivity
Conductivity measurement range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200 mS/cm
Conductivity accuracy ±0.5% F.S
Conductivity calibration points 1 to 5 points
Conductivity calibration solutions 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm or Custom
Conductivity resolution 0.001, 0.01, 0.1, 1
TDS
TDS range 0 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
TDS accuracy ±1 % F.S
TDS factor 0.01 to 1.00 (Default 0.5)
TDS resolution 0.01, 0.1, 1
Temperature
Temperature measuring range 0 105 °C, 32 to 221 °F
Temperature measuring accuracy ± °0.5 C, ± 0.9 °F
Temperature calibration points 1 point
Temperature resolution 0.1 °C, 0.1 °F
Temperature compensation 0 to 100 °C, 32 to 212 ° F, Manual or Automatic
Conductivity
Conductivity measurement range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200 mS/cm
Conductivity accuracy ±0.5% F.S
Conductivity calibration points 1 to 3 points
Conductivity calibration solutions 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm or Custom
Conductivity resolution 0.001, 0.01, 0.1, 1
Temperature
Temperature measuring range 0 105 °C, 32 to 221 °F
Temperature measuring accuracy ± °0.5 C, ± 0.9 °F
Temperature calibration points 1 point
Temperature resolution 0.1 °C, 0.1 °F
Temperature compensation 0 to 100 °C, 32 to 212 ° F, Manual or Automatic
Temperature coefficient Linear
Conductivity
Conductivity measurement range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200 mS/cm
Conductivity accuracy ±0.5% F.S
Conductivity calibration points 1 to 5 points
Conductivity calibration solutions 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm or Custom
Conductivity resolution 0.001, 0.01, 0.1, 1
Salinity
Salinity range 0.00 to 80.00 ppt, 0.00 to 42.00 psu
Salinity accuracy ±1% F.S
Salinity measurement modes Practical Salinity (psu) or Natural Seawater (ppt)
Salinity resolution 0.01, 0.1, 1
Temperature
Temperature measuring range 0 105 °C, 32 to 221 °F
Temperature measuring accuracy ± °0.5 C, ± 0.9 °F
Temperature calibration points 1 point
Temperature resolution 0.1 °C, 0.1 °F
Temperature compensation 0 to 100 °C, 32 to 212 ° F, Manual or Automatic
Measuring range Water activity 0 to 0.980 aw (no condensing )
Temperature range 0 to 50°C
Measurement accuracy ± 0.015 aw
Temperature accuracy ± 0.5°C
Range 0 to 1.0 aw
Resolution 0 to 1.0 aw
Accuracy ± 0.02 aw
Sensor Nonconductive humidity sensor