| pH Mode |
| pH range |
-2.00 to 20.00 pH |
| pH Accuracy |
± 0.002 pH |
| pH Resolution |
0.1/0.01/0.001 pH, selectable |
| pH Calibration Points |
1 to 5 Points |
| pH Buffer Solutions |
USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers |
|
| ORP Mode |
| mV Range |
± 2000 mV |
| mV Accuracy |
± 0.2 mV |
| mV Resolution |
0.1 mV, selectable |
| mV Calibration Points |
1 Point, only for relative mode |
|
| mV Mode |
|
| Ion Mode |
| Ion measurement range |
0.01 to 30000 ppm, mg/l, mol/l, mmol/l (refer Range of ion selection) |
| Ion measurement accuracy |
±0.5% F.S (Monovalent), ±1% F.S (Divalent) |
| Ion measurement resolution |
0.001/0.01/0.1/1 |
| Ion measurement calibration points |
2 to 5 points |
| Ion measurement calibration solutions |
0.001/0.01/0.1/1/10/100/1000/10000 ppm, mg/L, mol/L |
| Ion measurement methods |
Direct reading, Known addition, Known subtraction, Sample addition, Sample subtraction |
|
| Water Hardness Mode |
| Water hardness measurement range |
0.05-200mmol/L, 0-1120°dH, 0-1404°e, 0-2000°fH, 0-8000mg/L (Ca2+), 0~19999mg/L (CaCO3) |
| Water hardness measurement accuracy |
±1% F.S |
| Water hardness measurement resolution |
0.001/0.01/0.1/1 |
| Water hardness calibration points |
2 to 5 points |
| Water hardness calibration solutions |
0.01/0.1/1/10/100 mmol/L |
|
| Temperature Mode |
| Temperature measurement range |
0 to 105°, 32 to 212°F |
| Temperature measurement accuracy |
± 0.5°C, ± 0.9°F |
| Temperature measurement resolution |
0.1°C, 0.1°F |
| Temperature measurement calibration points |
1 point |
|
| Conductivity Mode |
|
| TDS Mode |
|
| Salinity Mode |
|
| Resistivity Mode |
|
| Dissolved Oxygen Mode |
|
| Temperature Range |
0 to 100°C, 32 to 221°F |
| Temperature Accuracy |
|
| Temperature Resolution |
|
| Temperature Calibrations Points |
|
| Temperature Compensation Mode |
Manual or Automatic |
| Data Hold Function |
Manual or Automatic end point detection |
| Stability Criteria |
Standard or high accuracy |
| Calibration Due Alarm |
1 to 31 days |
| Reset Function |
Yes |
| Memory |
Stores up to 1000 sets data |
| Output |
USB communication interface |
| Connector |
BNC Connector |
| Display |
7 inch TFT LCD display |
| Power supply |
DC 12 V, AC adapter, AC 220 V, 50 Hz |
| Dimension (L × W × H) |
240 × 220 × 80 mm |
| Weight |
1.7 kg |
- Common features
- Auto read feature to lock and read endpoint measurement
- Stability indicator to record stable measurement reading
- Reset feature automatically sets all function to default condition
- Automatic interval readings, sends the measurement data to the computer/printer
- USB communication interface to transfer store data
pH features
- 1-5 points push button calibration to automatically identify pH buffer
- Convenient to select pH buffer standard (USA/NIST/DIN) or custom calibration solutions
- Automatic temperature compensation, provides accurate readings over entire pH range
- Automatic electrode diagnosis with pH slope, offset display
- Alarm when pH readings are out of range
- ORP features
- 1 point offset calibration, can adjust display value to known standard
- Relative and absolute millivolt mode of measurement
- Water Hardness features 2 to 5 points calibration with high to low concentration
- Advanced measurements using different measurement units
- Ion features 2 to 5 2 to 5 points calibration, when 8 concentration points are selected
- Different concentration units like ppm, mg/l, mol/l, mmol/l for selection
- Automatic recognition of ion selective electrode
- Different Ion measurement methods for selection
- Calibration report with details of calibration standard and slope
| Accessories No. |
Name of the Ion electrode |
Range |
| 1 |
NH4 [Ammonium] Electrode |
5 × 10-6 ~1M |
| 2 |
Br [Bromide] Electrode |
5 × 10-6 ~1M |
| 3 |
Ca [Calcium] Electrode |
5 × 10-7 ~1M |
| 4 |
Cl [Chloride] Electrode |
5 × 10-6 ~1M |
| 5 |
Cn [Cyanide] Electrode |
5 × 10-6 ~0.01M |
| 6 |
F [Fluoride] Electrode |
1 × 10-6 M~Saturation |
| 7 |
NO3 [Nitrate] Electrode |
7 × 10-6 ~1M |
| 8 |
Ag [Silver] Electrode |
1 × 10-7 ~1M |
| 9 |
Na [Sodium[ Electrode |
1 × 10-5 ~1M |
| 10 |
Cd [Cadmium] Electrode |
1 × 10-6 ~0.1M |
| 11 |
Cu [Cupric] Electrode |
1 × 10-8 ~0.1M |
| 12 |
I [Iodide] Electrode |
5 × 10-8 ~1M |
| 13 |
Pb [Lead[ Electrode |
1 × 10-8 ~0.1M |
| 14 |
K [Potassium] Electrode |
1 × 10-6 ~1M |
| 15 |
S [Sulphide] Electrode |
1 × 10-7 ~1M |
| 16 |
NH3 [Ammonia] Electrode |
1 × 10-6 ~1M |
Used in multi-parameters based research testing , and quality analysis of water, other general liquids across water plants, food and beverages production units, laboratories, and institutes.