Benchtop Multi-parameter Water Quality Meter LMPWM-A40

Benchtop Multi-parameter Water Quality Meter LMPWM-A40 with 7 inch TFT display measures pH, mV, relative mV, water hardness, and temperature and Ion concentration of the sample. With expanded memory, it stores and recalls up to 1000 readings. Calibration due reminder informs operator to calibrate the meter regularly.

pH Mode
pH range -2.00 to 20.00 pH
pH Accuracy ± 0.002 pH
pH Resolution 0.1/0.01/0.001 pH, selectable
pH Calibration Points 1 to 5 Points
pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
ORP Mode
mV Range ± 2000 mV
mV Accuracy ± 0.2 mV
mV Resolution 0.1 mV, selectable
mV Calibration Points 1 Point, only for relative mode
mV Mode
Ion Mode
Ion measurement range 0.01 to 30000 ppm, mg/l, mol/l, mmol/l (refer Range of ion selection)
Ion measurement accuracy ±0.5% F.S (Monovalent), ±1% F.S (Divalent)
Ion measurement resolution 0.001/0.01/0.1/1
Ion measurement calibration points 2 to 5 points
Ion measurement calibration solutions 0.001/0.01/0.1/1/10/100/1000/10000 ppm, mg/L, mol/L
Ion measurement methods Direct reading, Known addition, Known subtraction, Sample addition, Sample subtraction
Water Hardness Mode
Water hardness measurement range 0.05-200mmol/L, 0-1120°dH, 0-1404°e, 0-2000°fH, 0-8000mg/L (Ca2+), 0~19999mg/L (CaCO3)
Water hardness measurement accuracy ±1% F.S
Water hardness measurement resolution 0.001/0.01/0.1/1
Water hardness calibration points 2 to 5 points
Water hardness calibration solutions 0.01/0.1/1/10/100 mmol/L
Temperature Mode
Temperature measurement range 0 to 105°, 32 to 212°F
Temperature measurement accuracy ± 0.5°C, ± 0.9°F
Temperature measurement resolution 0.1°C, 0.1°F
Temperature measurement calibration points 1 point
Conductivity Mode
TDS Mode
Salinity Mode
Resistivity Mode
Dissolved Oxygen Mode
Temperature Range 0 to 100°C, 32 to 221°F
Temperature Accuracy
Temperature Resolution
Temperature Calibrations Points
Temperature Compensation Mode Manual or Automatic
Data Hold Function Manual or Automatic end point detection
Stability Criteria Standard or high accuracy
Calibration Due Alarm 1 to 31 days
Reset Function Yes
Memory Stores up to 1000 sets data
Output USB communication interface
Connector BNC Connector
Display 7 inch TFT LCD display
Power supply DC 12 V, AC adapter, AC 220 V, 50 Hz
Dimension (L × W × H) 240 × 220 × 80 mm
Weight 1.7 kg
  • Common features
  • Auto read feature to lock and read endpoint measurement
  • Stability indicator to record stable measurement reading
  • Reset feature automatically sets all function to default condition
  • Automatic interval readings, sends the measurement data to the computer/printer
  • USB communication interface to transfer store data
    pH features
  • 1-5 points push button calibration to automatically identify pH buffer
  • Convenient to select pH buffer standard (USA/NIST/DIN) or custom calibration solutions
  • Automatic temperature compensation, provides accurate readings over entire pH range
  • Automatic electrode diagnosis with pH slope, offset display
  • Alarm when pH readings are out of range
  • ORP features
  • 1 point offset calibration, can adjust display value to known standard
  • Relative and absolute millivolt mode of measurement
  • Water Hardness features 2 to 5 points calibration with high to low concentration
  • Advanced measurements using different measurement units
  • Ion features 2 to 5 2 to 5 points calibration, when 8 concentration points are selected
  • Different concentration units like ppm, mg/l, mol/l, mmol/l for selection
  • Automatic recognition of ion selective electrode
  • Different Ion measurement methods for selection
  • Calibration report with details of calibration standard and slope
Accessories No. Name of the Ion electrode Range
1 NH4 [Ammonium] Electrode 5 × 10-6 ~1M
2 Br [Bromide] Electrode 5 × 10-6 ~1M
3 Ca [Calcium] Electrode 5 × 10-7 ~1M
4 Cl [Chloride] Electrode 5 × 10-6 ~1M
5 Cn [Cyanide] Electrode 5 × 10-6 ~0.01M
6 F [Fluoride] Electrode 1 × 10-6 M~Saturation
7 NO3 [Nitrate] Electrode 7 × 10-6 ~1M
8 Ag [Silver] Electrode 1 × 10-7 ~1M
9 Na [Sodium[ Electrode 1 × 10-5 ~1M
10 Cd [Cadmium] Electrode 1 × 10-6 ~0.1M
11 Cu [Cupric] Electrode 1 × 10-8 ~0.1M
12 I [Iodide] Electrode 5 × 10-8 ~1M
13 Pb [Lead[ Electrode 1 × 10-8 ~0.1M
14 K [Potassium] Electrode 1 × 10-6 ~1M
15 S [Sulphide] Electrode 1 × 10-7 ~1M
16 NH3 [Ammonia] Electrode 1 × 10-6 ~1M

Used in multi-parameters based research testing , and quality analysis of water, other general liquids across water plants, food and beverages production units, laboratories, and institutes.

pH Mode
pH range -2.00 to 20.00 pH
pH Accuracy ± 0.002 pH
pH Resolution 0.1/0.01/0.001 pH, selectable
pH Calibration Points 1 to 5 Points
pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
mV Mode
mV Range -1999.9 to 1999.9 mV
mV Accuracy ± 0.2 mV
mV Resolution 0.1/1 mV, selectable
mV Calibration Points 1 Point
Conductivity Mode
Conductivity Range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
Conductivity Accuracy ± 0.5% F.S.
Conductivity Resolution 0.01/0.1/1
Conductivity Calibration Points 1 to 5 Points
Conductivity Solutions 10, 84, 1413 µS/cm, 12.88, 111.8 mS/ cm
Conductivity Temperature coefficient 0.0 to 10.0 %/°C
Compensation mode Linear or Pure Water
Cell Constant K=0.1, 1, 10 or Custom
Reference temperature 20 or 25°C
Dissolved Oxygen Mode
Dissolved Oxygen Range 0.00 to 20.00 mg/ L
Dissolved Oxygen Accuracy ± 0.2 mg/ L
Dissolved Oxygen Resolution 0.1/ 0.01 mg/ L, selectable
Dissolved Oxygen Calibration Points 1 or 2 Points
% Saturation Range 0.0 to 200.0 % with ± 2 % accuracy
Pressure correction 60 to 112.5 kPa, 450 to 850 mmHg
Salinity correction 0 to 50 g/ L
pH Mode
pH range -2.00 to 20.00 pH
pH Accuracy ± 0.002 pH
pH Resolution 0.1/0.01/0.001 pH, selectable
pH Calibration Points 1 to 5 Points
pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
mV Mode
mV Range -1999.9 to 1999.9 mV
mV Accuracy ± 0.2 mV
mV Resolution 0.1/1 mV, selectable
mV Calibration Points 1 Point
Conductivity Mode
Conductivity Range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
Conductivity Accuracy ± 0.5% F.S.
Conductivity Resolution 0.01/0.1/1
Conductivity Calibration Points 1 to 5 Points
Conductivity Solutions 10, 84, 1413 µS/cm, 12.88, 111.8 mS/ cm
Conductivity Temperature coefficient 0.0 to 10.0 %/°C
Compensation mode Linear or Pure Water
Cell Constant K=0.1, 1, 10 or Custom
Reference temperature 20 or 25°C
TDS Mode
TDS Range 0 to 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
TDS Accuracy ± 1 % F.S.
TDS Factor 0.1  to 1.0 (default 0.5)
pH Mode
pH range -2.00 to 20.00 pH
pH Accuracy ± 0.002 pH
pH Resolution 0.1/0.01/0.001 pH, selectable
pH Calibration Points 1 to 5 Points
pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
mV Mode
mV Range -1999.9 to 1999.9 mV
mV Accuracy ± 0.2 mV
mV Resolution 0.1/1 mV, selectable
mV Calibration Points 1 Point
Conductivity Mode
Conductivity Range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
Conductivity Accuracy ± 0.5% F.S.
Conductivity Resolution 0.01/0.1/1
Conductivity Calibration Points 1 to 5 Points
Conductivity Solutions 10, 84, 1413 µS/cm, 12.88, 111.8 mS/ cm
Conductivity Temperature coefficient 0.0 to 10.0 %/°C
Compensation mode Linear or Pure Water
Cell Constant K=0.1, 1, 10 or Custom
Reference temperature 20 or 25°C
TDS Mode
TDS Range 0 to 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
TDS Accuracy ± 1 % F.S.
TDS Factor 0.1  to 1.0 (default 0.5)
Conductivity
Conductivity measurement range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200 mS/cm
Conductivity accuracy ±0.5% F.S
Conductivity calibration points 1 to 5 points
Conductivity calibration solutions 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm or Custom
Conductivity resolution 0.001, 0.01, 0.1, 1
Salinity
Salinity range 0.00 to 80.00 ppt, 0.00 to 42.00 psu
Salinity accuracy ±1% F.S
Salinity measurement modes Practical Salinity (psu) or Natural Seawater (ppt)
Salinity resolution 0.01, 0.1, 1
Temperature
Temperature measuring range 0 105 °C, 32 to 221 °F
Temperature measuring accuracy ± °0.5 C, ± 0.9 °F
Temperature calibration points 1 point
Temperature resolution 0.1 °C, 0.1 °F
Temperature compensation 0 to 100 °C, 32 to 212 ° F, Manual or Automatic
Conductivity
Conductivity measurement range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200 mS/cm
Conductivity accuracy ±0.5% F.S
Conductivity calibration points 1 to 3 points
Conductivity calibration solutions 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm or Custom
Conductivity resolution 0.001, 0.01, 0.1, 1
Temperature
Temperature measuring range 0 105 °C, 32 to 221 °F
Temperature measuring accuracy ± °0.5 C, ± 0.9 °F
Temperature calibration points 1 point
Temperature resolution 0.1 °C, 0.1 °F
Temperature compensation 0 to 100 °C, 32 to 212 ° F, Manual or Automatic
Temperature coefficient Linear
Conductivity
Conductivity measurement range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200 mS/cm
Conductivity accuracy ±0.5% F.S
Conductivity calibration points 1 to 5 points
Conductivity calibration solutions 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm or Custom
Conductivity resolution 0.001, 0.01, 0.1, 1
TDS
TDS range 0 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
TDS accuracy ±1 % F.S
TDS factor 0.01 to 1.00 (Default 0.5)
TDS resolution 0.01, 0.1, 1
Temperature
Temperature measuring range 0 105 °C, 32 to 221 °F
Temperature measuring accuracy ± °0.5 C, ± 0.9 °F
Temperature calibration points 1 point
Temperature resolution 0.1 °C, 0.1 °F
Temperature compensation 0 to 100 °C, 32 to 212 ° F, Manual or Automatic

Portable Water Hardness Meters

Labtron portable meters are used in quality analysis of water and other solutions. With portability, waterproof features, finds applications in indoor as well as outdoor measurement of pH, ORP, dissolved oxygen, water hardness, salinity, TDS etc.

Water hardness range 0.05 mmol / L - 0.2 mol / L<br/> (CaCO<sub>3</sub>) 0 - 19999 mg / L, (CaO) 0 - 11220 mg / L, (boiler) 0 - 400 mmol / L, (Ca) 0 - 8020 mg / L<br/> (German degree) 0 - 1122 °dH, (English degree) 0 - 1404 °e, (French degree) 0 - 2000 °f
Accuracy ± 1 % F.S.
Calibration points 2 to 5 points
Calibration solution 0.001, 0.01, 0.1 mmol / L, 0.001, 0.01, 0.1 mol / L

Benchtop Ion Meter

Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.

Ion measurement range F range: 1 × 10<sup>6</sup>M – saturation, 0.02 ppm to saturation
Ion measurement accuracy ± 0.5 % F.S.
Resolution 0.001, 0.01, 0.1, 1
Calibration points 2 to 5 points
Ion measurement range Cl range: 5 × 10<sup>6</sup> - 1 M, 1.8 - 35500 ppm
Ion measurement accuracy ± 0.5 % F.S.
Resolution 0.001, 0.01, 0.1, 1
Calibration points 2 to 5 points
Ion measurement range Na range: 1 × 10<sup>5</sup> - 1 M, 0.1 - 23000 ppm
Ion measurement accuracy ± 0.5 % F.S.
Resolution 0.001, 0.01, 0.1, 1
Calibration points 2 to 5 points
Measurement range 20000 mg/L <br/> 0.01 mg/L ~ 45000 mg/L <br/> 0.01 mg/L ~ 120000 mg/L
Measurement accuracy ± 2 % F.S. of its range ( Depending on the homogeneity of sludge)
Pressure range ≤ 0.4 Mpa
Flow rate ≤ 2.5 m/s, 8.2 ft./s

Salinity Tester

Labtron pocket testers are small and compact in size, gives fast and accurate results. The Pocket testers consist of pH, ORP, TDS, dissolved oxygen, salinity and conductivity tester for various laboratory as well as outdoor applications. Measurements can be taken by inserting the probe in to the test solution and after gentle stirring, stable readings can be noted down.

Salinity measurement range 0.00 to 80.00 ppt
Salinity accuracy ± 1 % F.S
Conductivity range 0 - 200.0 µS/cm, 100 - 2000 µS/cm, 20.00 mS/cm
Calibration solutions 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm
Salinity measurement range 0.00 to 10.00 ppt
Salinity accuracy ± 1 % F.S
Conductivity range 0 - 20.00, 200.0, 2000 µS/cm, 20.00 mS/cm
Calibration solutions 84 µS/cm, 1413 µS/cm, 12.88 mS/cm