Benchtop pH meter LBPM-A20

Benchtop pH meter LBPM-A20 is a high performance pH meter with large LCD display to read pH & temperature measurements. With -1.00 to 15.00 pH range it is suitable for daily measurement of pH in research laboratories as well as water quality testing units. Automatic calibration can be performed at 1-3 points using standard USA or NIST pH buffers. It offers automatic or manual temperature compensation of pH measurements.

pH measurement range -1.00 to 15.00 pH
pH accuracy ± 0.01 pH
pH resolution 0.01 pH
Calibration points 1-3 points
Calibration solutions USA (pH 4.0/7.00/10.01), NIST (pH 4.01/6.86/9.18)
ORP measurement range
ORP accuracy
ORP calibration points
ORP calibration range
ORP calibration solutions
Temperature compensation 0 to 100°C, 32 to 212°F
Compensation mode Manual or automatic
mV range -1000 to 1000 mV
mV accuracy ±1mV
mV resolution 1 mV
Temperature range 0 to 105°C, 32 to 221°F
Temperature accuracy ±1°C
Stability conditions
Calibration due
memory
Output
Temperature resolution 0.1°C
Hold function Manual or automatic endpoint detection
Power off Manual or automatic (180 minutes after last key is pressed)
Reset function Yes
Connector BNC
Display LCD
Dimension (L*W*H) 210×205×75 mm
Weight 1.5 kg
Power DC9V, AC adapter, AC220V/50Hz
  • Equipped with white backlit LCD display
  • pH measurement range: -1.00 to 15.00 pH
  • 1-3 points calibration to automatically identify standard pH buffer
  • Automatic electrode slope display helps operator whether to replace the sensor
  • Automatic temperature compensation for accurate measured value over the entire range
  • Auto hold function freezes stable, final readings for better viewing and recording
  • Manual temperature calibration to rectify the deviations in temperature
  • Universal BNC connector
  • Built-in solution holder
  • Reset function automatically resumes all the setting back to factory default options
  • Selectable temperature units (°C or °F) for different application requirements
  • Help message as an operational guide, helps users for quick access to the parameters
pH buffer solutions USA (pH 4.0/7.00/10.01), NIST (pH 4.01/6.86/9.18)
Temperature probe
pH probe suitable for various measurements (e.g. liquids, biofuels, Tris buffers, slurries, colloids, micro volume samples etc.)

Benchtop pH meter is used for research education purpose, quality testing in water and waste water management plants, food and beverage industry & other liquids & solvents.

Related Products of pH Meter (Benchtop)

Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.

pH measurement range -2.000 to 20.000 pH
pH accuracy ± 0.002 pH
Calibration points 1-5 points
Calibration solutions USA, NIST, DIN or user defined
pH measurement range 0.00 to 14.00 pH
pH accuracy ± 0.01 pH
pH resolution 0.01 pH
Calibration points 2 points
Temperature compensation 0 to 100°C, 32°F to 212°F Manual or Automatic
Hold function Manual or automatic endpoint detection
Stability criteria Standard or High-accuracy
Interval readings 10, 30, 60 seconds, 10, 30 minutes or Off
Determination range
0 to 35 mg/L, 0 to 70 mg/ L, 0 to 350 mg/L, 0 to 700 mg/L (without dilution)
0 to 4000 mg/ L (need to dilute) (subsection measurement)
Measurement error Less or less than 10 %
Sample container/ bottle quantity 1 to 6
Sample container/ bottle Capacity 1000 ml

Salinity Tester

Labtron pocket testers are small and compact in size, gives fast and accurate results. The Pocket testers consist of pH, ORP, TDS, dissolved oxygen, salinity and conductivity tester for various laboratory as well as outdoor applications. Measurements can be taken by inserting the probe in to the test solution and after gentle stirring, stable readings can be noted down.

Salinity measurement range 0.00 to 10.00 ppt
Salinity accuracy ± 1 % F.S
Conductivity range 0 - 20.00, 200.0, 2000 µS/cm, 20.00 mS/cm
Calibration solutions 84 µS/cm, 1413 µS/cm, 12.88 mS/cm
Salinity measurement range 0.00 to 80.00 ppt
Salinity accuracy ± 1 % F.S
Conductivity range 0 - 200.0 µS/cm, 100 - 2000 µS/cm, 20.00 mS/cm
Calibration solutions 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm

Conductivity Tester

Labtron pocket testers are small and compact in size, gives fast and accurate results. The Pocket testers consist of pH, ORP, TDS, dissolved oxygen, salinity and conductivity tester for various laboratory as well as outdoor applications. Measurements can be taken by inserting the probe in to the test solution and after gentle stirring, stable readings can be noted down.

Conductivity measurement range 1.0 to 199.9 µS/cm
Type of conductivity Low range conductivity
Conductivity accuracy ± 1 % F.S.
Conductivity calibration points 1 point
Conductivity measurement range 10 to 1999 µS/cm
Type of conductivity Mid-range conductivity
Conductivity accuracy ± 1 % F.S.
Conductivity calibration points 1 point
Conductivity measurement range 0.1 to 19.99 mS/cm
Type of conductivity High range conductivity
Conductivity accuracy ± 1 % F.S.
Conductivity calibration points 1 point
COD
COD determination range 2 to 10000 mg/ L
COD measurement accuracy COD < 50 mg/ L, ≤ ± 10 % and COD > 50 mg/ L, ≤ ± 5 %
COD minimum detection limit 0.1 mg/ L
COD measurement time 20 minutes
COD repeatability ≤ ± 5 %
Processing sample number per batch
Consumables for colorimetric COD measurement Colorimetric tubes and Cuvette
COD reactor digestion temperature 165 ± 0.5°C
COD reactor digestion time 10 min
COD reactor time setting switches 3 pcs
Anti-chlorine interference (Cl) < 1000 mg/ L No effect
Ammonia nitrogen
Ammonia nitrogen determination range 0.02 to 60 mg/ L
Ammonia nitrogen measurement accuracy ≤ ± 10 %
Ammonia nitrogen minimum detection limit 0.01 mg/ L
Ammonia nitrogen measurement time 10 to 15 minutes
Ammonia nitrogen repeatability ≤ ± 5 %
Processing sample number per batch
Consumables for colorimetric ammonia nitrogen measurement Colorimetric tubes and Cuvette
Ammonia nitrogen reactor time setting switches 3 pcs
Total phosphorous
Total phosphorous determination range 0.002 to 20 mg/ L
Total phosphorous total phosphorous measurement accuracy ≤ ± 5 %
Total phosphorous minimum detection limit 0.001 mg/ L
Total phosphorous measurement time 35 to 50 minutes
Total phosphorous repeatability ≤ ± 5 %
Processing sample number per batch
Consumables for colorimetric total phosphorous measurement Colorimetric tubes and Cuvette
Total phosphorous reactor digestion temperature 120 ± 0.5°C
Total phosphorous reactor digestion time 30 min
Total phosphorous reactor time setting switches 3 pcs
Turbidity
Turbidity determination range 0.5 to 400 NTU
Turbidity measurement accuracy ≤ ± 5 %
Turbidity minimum detection limit 0.1 NTU
Turbidity measurement time 1 minute
Turbidity repeatability ≤ ± 5 %
Consumables for colorimetric turbidity measurement Colorimetric Cuvette
COD
COD determination range 5 to 5000 mg/ L
COD measurement accuracy COD < 50 mg/ L, ≤ ± 10 % and COD > 50 mg/ L, ≤ ± 5 %
COD minimum detection limit 0.1 mg/ L
COD measurement time 20 minutes
COD repeatability ≤ ± 5 %
Processing sample number per batch 12 pcs
Consumables for colorimetric COD measurement Colorimetric tubes and Cuvette
COD reactor digestion temperature 165 ± 0.5°C
COD reactor digestion time 10 min
COD reactor time setting switches 3 pcs
Anti-chlorine interference (Cl) < 1000 mg/ L No effect
Ammonia nitrogen
Ammonia nitrogen determination range 0.02 to 30 mg/ L
Ammonia nitrogen measurement accuracy ≤ ± 10 %
Ammonia nitrogen minimum detection limit 0.01 mg/ L
Ammonia nitrogen measurement time 10 to 15 minutes
Ammonia nitrogen repeatability ≤ ± 5 %
Processing sample number per batch 20 pcs
Consumables for colorimetric ammonia nitrogen measurement Colorimetric tubes and Cuvette
Ammonia nitrogen reactor time setting switches 3 pcs
Total phosphorous
Total phosphorous determination range 0.002 to 10 mg/ L
Total phosphorous total phosphorous measurement accuracy ≤ ± 5 %
Total phosphorous minimum detection limit 0.001 mg/ L
Total phosphorous measurement time 35 to 50 minutes
Total phosphorous repeatability ≤ ± 5 %
Processing sample number per batch 12 pcs
Consumables for colorimetric total phosphorous measurement Colorimetric tubes and Cuvette
Total phosphorous reactor digestion temperature 120 ± 0.5°C
Total phosphorous reactor digestion time 30 min
Total phosphorous reactor time setting switches 3 pcs
Turbidity
Turbidity determination range 0.5 to 400 NTU
Turbidity measurement accuracy ≤ ± 5 %
Turbidity minimum detection limit 0.1 NTU
Turbidity measurement time 1 minute
Turbidity repeatability ≤ ± 5 %
Consumables for colorimetric turbidity measurement Colorimetric Cuvette
pH Mode
pH range -2.00 to 20.00 pH
pH Accuracy ± 0.002 pH
pH Resolution 0.1/0.01/0.001 pH, selectable
pH Calibration Points 1 to 5 Points
pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
mV Mode
mV Range -1999.9 to 1999.9 mV
mV Accuracy ± 0.2 mV
mV Resolution 0.1/1 mV, selectable
mV Calibration Points 1 Point
Conductivity Mode
Conductivity Range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
Conductivity Accuracy ± 0.5% F.S.
Conductivity Resolution 0.01/0.1/1
Conductivity Calibration Points 1 to 5 Points
Conductivity Solutions 10, 84, 1413 µS/cm, 12.88, 111.8 mS/ cm
Conductivity Temperature coefficient 0.0 to 10.0 %/°C
Compensation mode Linear or Pure Water
Cell Constant K=0.1, 1, 10 or Custom
Reference temperature 20 or 25°C
Dissolved Oxygen Mode
Dissolved Oxygen Range 0.00 to 20.00 mg/ L
Dissolved Oxygen Accuracy ± 0.2 mg/ L
Dissolved Oxygen Resolution 0.1/ 0.01 mg/ L, selectable
Dissolved Oxygen Calibration Points 1 or 2 Points
% Saturation Range 0.0 to 200.0 % with ± 2 % accuracy
Pressure correction 60 to 112.5 kPa, 450 to 850 mmHg
Salinity correction 0 to 50 g/ L
pH Mode
pH range -2.00 to 20.00 pH
pH Accuracy ± 0.002 pH
pH Resolution 0.1/0.01/0.001 pH, selectable
pH Calibration Points 1 to 5 Points
pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
mV Mode
mV Range -1999.9 to 1999.9 mV
mV Accuracy ± 0.2 mV
mV Resolution 0.1/1 mV, selectable
mV Calibration Points 1 Point
Conductivity Mode
Conductivity Range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
Conductivity Accuracy ± 0.5% F.S.
Conductivity Resolution 0.01/0.1/1
Conductivity Calibration Points 1 to 5 Points
Conductivity Solutions 10, 84, 1413 µS/cm, 12.88, 111.8 mS/ cm
Conductivity Temperature coefficient 0.0 to 10.0 %/°C
Compensation mode Linear or Pure Water
Cell Constant K=0.1, 1, 10 or Custom
Reference temperature 20 or 25°C
TDS Mode
TDS Range 0 to 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
TDS Accuracy ± 1 % F.S.
TDS Factor 0.1  to 1.0 (default 0.5)
pH Mode
pH range -2.00 to 20.00 pH
pH Accuracy ± 0.002 pH
pH Resolution 0.1/0.01/0.001 pH, selectable
pH Calibration Points 1 to 5 Points
pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
mV Mode
mV Range -1999.9 to 1999.9 mV
mV Accuracy ± 0.2 mV
mV Resolution 0.1/1 mV, selectable
mV Calibration Points 1 Point
Conductivity Mode
Conductivity Range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
Conductivity Accuracy ± 0.5% F.S.
Conductivity Resolution 0.01/0.1/1
Conductivity Calibration Points 1 to 5 Points
Conductivity Solutions 10, 84, 1413 µS/cm, 12.88, 111.8 mS/ cm
Conductivity Temperature coefficient 0.0 to 10.0 %/°C
Compensation mode Linear or Pure Water
Cell Constant K=0.1, 1, 10 or Custom
Reference temperature 20 or 25°C
TDS Mode
TDS Range 0 to 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
TDS Accuracy ± 1 % F.S.
TDS Factor 0.1  to 1.0 (default 0.5)