Ambient Temperature |
10°C to 35°C |
Measurement Range |
0 to 100 mg |
Resolution |
0.1 µg |
Max. Electrolysis Speed |
40 µg /s |
Amount and Precision of electrolysed water |
10 µg to 1000 µg, ± 2 µg and >1000 µg, 0.2 % |
Relative Humidity |
≤ 85 % |
Max. Power Consumption |
30 W |
Power Supply |
AC 220 V ± 20 %, 50 Hz |
Dimension |
320 × 240 × 150 mm |
Net Weight |
5 kg |
- Capability of saving maximum 255 data
- 240 × 128 pixels dot-matrix LCD screen
- Equipped with Calendar clock, to record date and time
- End point of sample is measured with high electrode precision value with electrolyte blank current and equilibrium point drift compensation adaptation for accurate measurement
- Automatic shutdown LCD screen with screen protection function
- Design with open-circuit fault and short-circuit fault for automatic detecting function of measuring electrode
- Portable and easy handling
Coulometric Karl Fischer Titrator is used for the determination of water content in petroleum products and lubricating oils and additives by Karl Fischer titration